首页> 外文期刊>Arthroscopy: the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association >Three femoral fixation devices for anterior cruciate ligament reconstruction: comparison of fixation on the lateral cortex versus the anterior cortex.
【24h】

Three femoral fixation devices for anterior cruciate ligament reconstruction: comparison of fixation on the lateral cortex versus the anterior cortex.

机译:三种用于前交叉韧带重建的股骨固定装置:外侧皮质与前皮质固定的比较。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: To evaluate the biomechanical properties of 3 anterior cruciate ligament (ACL) reconstruction femoral fixation devices in a porcine model with implantation on both the lateral femoral cortex and the anterior femoral cortex. METHODS: ACL reconstructions with an 8-mm porcine tendon graft were performed on 48 porcine femurs with the EndoButton CL (Smith & Nephew, Andover, MA), ToggleLoc with ZipLoop technology (Biomet Sports Medicine, Warsaw, IN), or EZLoc (Biomet Sports Medicine). In 8 specimens for each implant, the femoral tunnel was drilled from the 10:30 surgical position out the lateral cortex. In another 8 specimens for each implant, the tunnel was drilled from the 10:30 position to a standardized anterior femoral surface. Cyclic testing was performed on an MTS testing machine (MTS, Eden Prairie, MN) from 50 N to 450 N for 2,000 cycles, followed by load-to-failure testing in specimens that survived. The cortical thickness and location of the implant exit were recorded. RESULTS: In the lateral femur group, 0 of the EZLoc devices, 2 of the ToggleLoc devices, and 3 of the EndoButtons completed cyclic testing. In the anterior femur group, 1 of the EZLoc devices, 5 of the ToggleLoc devices, and 5 of the EndoButtons completed cyclic testing (P = .012). In the anterior femur group, the ToggleLoc exhibited higher 2,000-cycle elongation (5.46 +/- 1 mm) than the EndoButton (3.55 +/- 0.6 mm) (P = .005). The EndoButton showed a higher first failure load (1,190.9 +/- 150.0 N) than the ToggleLoc (912.6 +/- 82.4 N) (P = .007). The anterior cortex (1.4 mm) was thinner than the lateral cortex (1.7 mm) (P = .0002). CONCLUSIONS: The EndoButton provided the strongest ACL femoral fixation with significantly less graft-implant elongation and significantly higher failure loads. It was also shown in a porcine model that implants on the anterior cortical surface perform better than those on the lateral surface. Increased cortical thickness, in the range tested, was not associated with improved implant performance in the porcine model. CLINICAL RELEVANCE: The EndoButton provided the best ACL femoral fixation of the devices tested.
机译:目的:在猪模型中评估3个前交叉韧带(ACL)重建股骨固定装置的生物力学特性,并在股外侧皮和股骨前皮上植入。方法:采用EndoButton CL(Smith&Nephew,Andover,MA),采用ZipLoop技术的ToggleLoc(Biomet Sports Medicine,Warsaw,IN)或EZLoc(Biomet)对8头猪股骨进行8毫米猪腱移植的ACL重建。运动药物)。对于每个植入物,在8个标本中,从10:30手术位置在外侧皮层钻出股骨隧道。在每个植入物的另外8个标本中,从10:30的位置钻至标准的股前表面。在50 N至450 N的MTS测试机(MTS,Eden Prairie,MN)上进行了2,000次循环测试,然后对幸存的样品进行了载荷-破坏测试。记录皮质厚度和植入物出口的位置。结果:在股骨外侧组中,0台EZLoc设备,2台ToggleLoc设备和3台EndoButtons完成了循环测试。在股骨前组中,有1个EZLoc设备,5个ToggleLoc设备和5个EndoButtons完成了循环测试(P = .012)。在股骨前组中,ToggleLoc的2,000周期伸长率(5.46 +/- 1 mm)比EndoButton(3.55 +/- 0.6 mm)高(P = .005)。与ToggleLoc(912.6 +/- 82.4 N)(P = .007)相比,EndoButton显示出更高的首次故障负荷(1,190.9 +/- 150.0 N)。前皮质(1.4 mm)比侧皮质(1.7 mm)薄(P = .0002)。结论:EndoButton提供了最强的ACL股骨固定,移植物-植入物的伸长率明显降低,而失败负荷明显升高。在猪模型中还表明,前皮质表面的植入物比侧表面的植入物性能更好。在测试范围内,增加的皮层厚度与猪模型中植入物性能的改善无关。临床相关性:EndoButton对所测试的设备提供了最好的ACL股骨固定。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号