Sudden power loss in a heartmate II left ventricular assist device due to intermittent pin contact with the battery: Case report

机译:间歇性引脚与电池接触导致心室II左心室辅助设备突然断电:病例报告

获取原文
获取原文并翻译 | 示例
           

摘要

Left ventricular assist device technology has improved such that mechanical malfunction, particularly with newer generation continuous flow devices, is a relatively rare event. We present a case of sudden power loss in a HeartMate II caused by intermittent contact of the battery terminals after a clip was dropped with the battery inserted in it. The clip was replaced and the patient made a complete recovery. A new inspection and testing methodology, and amended approach to patient and caregiver training, designed to prevent future occurrences is described.
机译:左心室辅助设备技术已得到改进,使得机械故障(尤其是新一代的连续流设备)相对少见。我们介绍了一种在HeartMate II中突然断电的情况,该断电是由于在将夹子插入电池后掉落夹子而导致电池端子间断接触引起的。夹子被更换,患者完全康复。描述了一种旨在防止将来发生的新的检查和测试方法,以及对患者和护理人员进行培训的修订方法。