The design and initial test results of a new passively suspended Tesla type left ventricular assist device blood pump are described. Computational fluid dynamics (CFD) analysis was used in the design of the pump. Overall size of the prototype device is 50 mm in diameter and 75 mm in length. The pump rotor has a density lower than that of blood and when spinning inside the stator in blood it creates a buoyant centering force that suspends the rotor in the radial direction. The axial magnetic force between the rotor and stator restrain the rotor in the axial direction. The pump is capable of pumping up to 10 L/min at a 70 mm Hg head rise at 8,000 revolutions per minute (RPM). The pump has demonstrated a normalized index of hemolysis level below 0.02 mg/dL for flows between 2 and 9.7 L/min. An inlet pressure sensor has also been incorporated into the inlet cannula wall and will be used for control purposes. One initial in vivo study showed an encouraging result. Further CFD modeling refinements are planned and endurance testing of the device.
机译:描述了新型被动悬挂的特斯拉型左心室辅助设备血泵的设计和初步测试结果。在泵的设计中使用了计算流体动力学(CFD)分析。原型设备的总尺寸为直径50毫米,长度75毫米。泵转子的密度低于血液的密度,当定子在血液中在定子内旋转时,会产生浮力的定心力,该定心力使转子沿径向悬挂。转子和定子之间的轴向磁力沿轴向限制转子。该泵能够以8,000转/分钟(RPM)的速度以70毫米汞柱的压头上升速度泵送高达10 L / min的流量。对于2至9.7 L / min的流量,该泵已显示出低于0.02 mg / dL的正常溶血指数。入口压力传感器也已合并到入口套管壁中,并将用于控制目的。一项初步的体内研究显示了令人鼓舞的结果。计划进一步完善CFD建模,并对设备进行耐久性测试。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号