首页> 外文期刊>Bone >Specimen diameter and 'side artifacts' in cancellous bone evaluated using end-constrained elastic tension.
【24h】

Specimen diameter and 'side artifacts' in cancellous bone evaluated using end-constrained elastic tension.

机译:使用末端约束的弹性张力评估松质骨中的标本直径和“侧面伪影”。

获取原文
获取原文并翻译 | 示例
           

摘要

In cancellous bone testing of cored samples, side artifacts are the underestimation of the true (i.e. in situ) mechanical properties due to the severing of the trabecular network during specimen preparation. Although other researchers have suggested correction factors derived from finite element method (FEM) models, it is proposed that side effects can be minimized by increasing the specimen diameter. Six different diameter specimens (3.1-10.6 mm), from two different anatomic sites (bovine femoral condyle and bovine lumbar vertebrae), were mechanically tested in elastic tension using an epoxy endcap protocol to eliminate end artifacts. Elastic modulus was found to be significantly affected by diameter in both sites. For example, the 5.1 mm samples underestimated the elastic modulus of the 10.6 mm samples by an average of roughly 20%. Yet no statistical difference was detected between the 8.3 and 10.6 mm samples in either anatomic site, suggesting that 8.3 mm diameter specimens were sufficiently large to avoid side artifacts. FEM models created from micro-CT images reveal that modulus approaches an asymptotic value with increasing diameter, and demonstrate an architecture-dependent drop in modulus at decreasing diameters. These results confirm, both experimentally and numerically, that side effects can be ignored given a suitably large specimen diameter and that this minimum diameter will be dependent on the cancellous architecture. An important implication of the latter result is that specimen diameters must be chosen appropriately when comparing test groups with different architectures (e.g. normal versus osteoporotic) to ensure that the magnitude of side artifacts does not confound the true differences between the groups.
机译:在有芯样品的松质骨测试中,由于在准备样品过程中切断了小梁网络,侧面伪影低估了真实(即原位)机械性能。尽管其他研究人员已经提出了源自有限元方法(FEM)模型的校正因子,但建议可以通过增加样品直径来最大程度地减少副作用。对来自两个不同解剖部位(牛股骨con和牛腰椎)的六个不同直径的标本(3.1-10.6 mm)进行了机械测试,采用环氧端帽协议以消除末端伪影,进行了弹性拉伸。发现弹性模量在两个部位均受直径显着影响。例如,5.1毫米样品低估了10.6毫米样品的弹性模量,平均低约20%。然而,在任何一个解剖部位的8.3 mm和10.6 mm样品之间均未检测到统计学差异,这表明8.3 mm直径的样品足够大,可以避免出现侧面伪影。从微型CT图像创建的FEM模型显示,模量随着直径的增加而接近渐近值,并证明了随着直径的减小,模量在体系结构方面的下降。这些结果在实验和数值上都证实,给定适当大的样品直径,可以忽略副作用,并且该最小直径将取决于松质结构。后一个结果的重要含义是,在比较具有不同结构(例如正常与骨质疏松)的测试组时,必须适当选择样本直径,以确保侧面伪影的大小不会混淆各组之间的真实差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号