Minimizing specimen length in elastic testing of end-constrained cancellous bone.

机译:在末端约束松质骨的弹性测试中最小化样本长度。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

End-constraint is needed when performing tensile and compressive testing of cancellous bone so that frictional and structural end-effects are minimized and the mechanical properties are not underestimated. Since many end-constrained testing techniques are limited to certain sites and species by the large specimen sizes required, reductions in overall specimen length would be advantageous. The following study examines the effect of specimen gauge length and aspect ratio on the elastic tensile testing of cancellous bone from the bovine femoral condyle. Three different nominal gauge lengths (10, 16, 22 mm) were examined at two different diameters (5.1, 10.7 mm) which bound the range typically used for cancellous bone testing. No effect of gauge length or aspect ratio was observed, but evidence of a diameter effect was noted. Three-dimensional finite element method (FEM) models confirmed the experimental observations and demonstrated that an overestimation of modulus begins to occur for specimens shorter than 10 mm.
机译:在进行松质骨的拉伸和压缩测试时,需要进行最终约束,以使摩擦和结构的最终影响最小化,并且不会低估机械性能。由于所需的大样本尺寸限制了许多最终受限的测试技术,因此仅限于某些地点和物种,因此减少整体样本长度将是有利的。以下研究检查了标距长度和纵横比对牛股骨con松质骨弹性拉伸测试的影响。在两个不同的直径(5.1、10.7 mm)下检查了三种不同的标称标距长度(10、16、22 mm),这些直径限制了通常用于松质骨测试的范围。没有观察到标距长度或纵横比的影响,但是注意到了直径影响的证据。三维有限元方法(FEM)模型证实了实验观察结果,并证明对于小于10 mm的样品,模量开始出现高估。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号