Accuracy of Microlife WatchBP Office ABI monitor assessed according to the 2002 European Society of Hypertension protocol and the British Hypertension Society protocol.

机译:Microlife WatchBP Office ABI监测仪的准确性已根据2002年欧洲高血压学会协议和英国高血压学会协议进行了评估。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: To determine the accuracy of the WatchBP Office ABI monitor for blood pressure measurement developed by the Microlife Company. METHODS: The device accuracy was tested in 85 subjects with a mean age of 54 +/- 19 years. Their systolic and diastolic blood pressure (SBP/DBP) at entry was 141 +/- 30/86 +/- 19 mmHg, and upper arm circumference was 28 +/- 5 cm. Initially, the data from 33 participants were examined according to the 2002 version of the European Society of Hypertension (ESH) protocol. An additional 52 subjects were then enrolled to fulfill the requirements of the British Hypertension Society (BHS) protocol. In all participants, sequential same arm measurements were performed by two trained observers. RESULTS: The device passed all three phases of the ESH protocol for SBP and DBP. For the BHS protocol the device was graded A for both SBP and DBP. The A/A grade was achieved in the low blood pressure category (< 130/80 mmHg), the B/A grade in the medium category (130-160/80-100 mmHg) and the A/A grade in the high category (> 160/100 mmHg). Mean blood pressure difference between device and observers in the first 33 subjects was -0.9 +/- 5.5 mmHg for SBP and -2.2 +/- 4.5 mmHg for DBP and in the 85 participants it was -1.2 +/- 6.5 mmHg and -2.3 +/- 5.1, respectively. CONCLUSION: These data show that the Microlife WatchBP Office ABI monitor satisfied the recommended ESH accuracy levels and achieved A/A grade of the BHS protocol across a wide range of BP.
机译:目的:确定由Microlife公司开发的WatchBP Office ABI血压测量仪的准确性。方法:在85位平均年龄为54 +/- 19岁的受试者中测试了设备的准确性。他们进入时的收缩压和舒张压(SBP / DBP)为141 +/- 30/86 +/- 19 mmHg,上臂围为28 +/- 5 cm。最初,根据2002年欧洲高血压学会(ESH)协议检查了33位参与者的数据。然后又招募了52名受试者,以符合英国高血压学会(BHS)协议的要求。在所有参与者中,由两名训练有素的观察者进行了顺序相同的手臂测量。结果:该设备通过了SBP和DBP的ESH协议的所有三个阶段。对于BHS协议,该设备的SBP和DBP均为A级。低血压类别(<130/80 mmHg)达到A / A等级,中类别(130-160 / 80-100 mmHg)达到B / A等级,高类别达到A / A等级(> 160/100 mmHg)。在前33名受试者中,设备与观察者之间的平均血压差对于SBP为-0.9 +/- 5.5 mmHg,对于DBP为-2.2 +/- 4.5 mmHg,在85位参与者中,其为-1.2 +/- 6.5 mmHg和-2.3分别为+/- 5.1。结论:这些数据表明,Microlife WatchBP Office ABI监护仪满足了推荐的ESH准确性水平,并在广泛的BP范围内达到了BHS协议的A / A级。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号