Validation of Heine Gamma G7 (G5) and XXL-LF aneroid devices for blood pressure measurement.

机译:验证Heine Gamma G7(G5)和XXL-LF无液装置用于血压测量。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: To determine the accuracy of the Gamma G7 (and G5 model) and XXL-LF aneroid sphygmomanometers developed by the Heine Company. DESIGN: Device evaluations were performed using the new protocol of the European Society of Hypertension. Monitor performance was assessed in relation to participants' sex, age, arm circumference, and systolic and diastolic blood pressures. METHODS: The two sphygmomanometers were assessed in two different samples according to European Society of Hypertension requirements, which are based on four zones of accuracy differing from the mercury standard by 5, 10, 15 mmHg, or more. RESULTS: Both sphygmomanometers passed all three phases of the protocol for systolic blood pressure and diastolic blood pressure. Mean blood pressure difference between Gamma G7 sphygmomanometer and observers was -0.4+/-3.3 mmHg for systolic blood pressure and -0.5+/-2.6 mmHg for diastolic blood pressure. Mean differences for the Gamma XXL-LF sphygmomanometer were -0.3+/-3.7 and -1.0+/-2.6 mmHg, respectively. In multivariable analyses, the SBP discrepancies between both aneroid sphygmomanometers and observers were unrelated to age, sex, arm circumference and systolic blood pressure. For diastolic blood pressure, a borderline relationship was found only for arm circumference (P=0.057) with the Gamma G7 device. CONCLUSIONS: These data show that the Heine Gamma G7 and Gamma XXL-LF aneroid sphygmomanometers satisfy the new recommended ESH accuracy levels for both SBP and DBP. Their performance is uniform across subgroups of participants with different clinical characteristics.
机译:目的:确定Heine公司开发的Gamma G7(和G5型号)和XXL-LF无液血压计的准确性。设计:使用欧洲高血压学会的新方案进行设备评估。根据参与者的性别,年龄,臂围以及收缩压和舒张压评估监护仪的性能。方法:根据欧洲高血压学会的要求,在两个不同的样本中对这两个血压计进行了评估,这是基于与汞标准相差5、10、15 mmHg或更高的四个精度区域。结果:两种血压计均通过了协议的所有三个阶段的收缩压和舒张压。 Gamma G7血压计和观察者之间的平均血压差为收缩压为-0.4 +/- 3.3 mmHg,而舒张压为-0.5 +/- 2.6 mmHg。 Gamma XXL-LF血压计的平均差分别为-0.3 +/- 3.7和-1.0 +/- 2.6 mmHg。在多变量分析中,无液血压计和观察者之间的SBP差异与年龄,性别,臂围和收缩压无关。对于舒张压,仅在使用Gamma G7装置的手臂周围(P = 0.057)发现了临界线关系。结论:这些数据表明,Heine Gamma G7和Gamma XXL-LF无液血压计满足SBP和DBP的新推荐ESH准确度水平。在具有不同临床特征的参与者亚组中,他们的表现是统一的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号