首页> 外文期刊>Blood Pressure Monitoring >Validation of the Kingyield BP210 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension-International Protocol
【24h】

Validation of the Kingyield BP210 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension-International Protocol

机译:根据欧洲高血压学会-国际协议对Kingyield BP210腕式血压监测仪进行家庭血压监测的验证

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: The present study aimed to evaluate the accuracy of the automated oscillometric wrist blood pressure monitor BP210 for home blood pressure monitoring according to the International Protocol of the European Society of Hypertension. METHODS: Systolic and diastolic blood pressures were sequentially measured in 33 adult Chinese participants (21 women, 51 years of mean age) using a mercury sphygmomanometer (two observers) and the BP210 device (one supervisor). Ninety-nine pairs of comparisons were obtained from 15 participants in phase 1 and a further 18 participants in phase 2 of the validation study. Data analysis was conducted using the ESHIP analyzer. RESULTS: The BP210 device successfully passed phase 1 of the validation study with a number of absolute differences between device and observers within 5, 10, and 15 mmHg for at least 33/45, 44/45, and 44/45 measurements, respectively. The device also achieved the targets for phase 2.1, with 77/99, 95/99, and 97/99 differences within 5, 10, and 15 mmHg, respectively for systolic blood pressure, and with 78/99, 97/99, and 99/99 within 5, 10, and 15 mmHg, respectively for diastolic blood pressure. In phase 2.2, 29 and 25 participants had at least two of the three device-observers differences within 5 mmHg (required ≥22) for systolic blood pressure and diastolic blood pressure, respectively. CONCLUSION: The Kingyield wrist blood pressure monitor BP210 has passed the International Protocol requirements, and hence can be recommended for home use in adults.
机译:目的:本研究旨在评估根据欧洲高血压学会国际协议,用于家庭血压监测的自动示波腕式血压计BP210的准确性。方法:使用水银血压计(两名观察员)和BP210装置(一名主管)对33名中国成年参与者(21名妇女,平均年龄51岁)依次测量收缩压和舒张压。从验证研究的第一阶段的15名参与者和第二阶段的另外18名参与者中获得了99对比较。使用ESHIP分析仪进行数据分析。结果:BP210设备成功通过了验证研究的第1阶段,在至少33 / 45、44 / 45和44/45的测量中,设备与观察者之间的绝对差值分别在5、10和15 mmHg之内。该设备还达到了2.1阶段的目标,收缩压的差异分别在5、10和15 mmHg之内,分别为77 / 99、95 / 99和97/99,而78 / 99、97 / 99和97舒张压分别在5、10和15 mmHg之内达到99/99。在阶段2.2中,有29位参与者和25位参与者在收缩压和舒张压分别在5 mmHg(要求≥22)内的三个设备观察者差异中至少有两个。结论:Kingyield腕式血压计BP210已通过国际协议的要求,因此可建议成人使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号