Repeatability of Ocular Measurements with a Dual-Scheimpflug Analyzer in Healthy Eyes

机译:双Scheimpflug分析仪在健康眼睛中眼图测量的可重复性

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Purpose. To assess the repeatability of the Galilei dual Scheimpflug analyzer (GDSA) in anterior segment examination. Methods. Fifty-two eyes from 52 healthy volunteers were prospectively and consecutively recruited. Anatomic, axial, refractive, and instantaneous parameters were measured with GDSA to provide a complete characterization of the anterior segment. Repeatability was assessed calculating intraclass correlation coefficient (ICC), and coefficient of variation (COV). Results. Correlation among repeated measurements showed almost perfect reliability (ICC > 0.81) for all parameters except thinnest central corneal thickness (CCT) (0.78), corneal thickness average out (0.79), and posterior axial curvature average out (0.60). Repeatability was excellent (COV < 10%) for all parameters except anterior chamber volume and, superior iridocorneal angle and eccentricities. In these last three parameters, repeatability limits were excessively high compared to the mean. Conclusions. GDSA in healthy young persons had an almost perfect correlation in measuring anatomic, axial, instantaneous, and refractive parameters with greater variability for peripheral terms. Repeatability of anatomical parameters like pachymetry, anterior chamber, or iridocorneal angle and eccentricity were limited. In healthy young persons, the other evaluated parameters had very good repeatability and their limits of agreement showed excellent clinical results for this device.
机译:目的。为了评估前节检查中Galilei双Scheimpflug分析仪(GDSA)的可重复性。方法。来自52名健康志愿者的52只眼睛被前瞻性和连续地招募。用GDSA测量解剖,轴向,屈光和瞬时参数,以提供前节的完整特征。通过计算组内相关系数(ICC)和变异系数(COV)评估重复性。结果。重复测量之间的相关性显示,除最薄的中央角膜厚度(CCT)(0.78),角膜厚度平均(0.79)和后轴曲率平均(0.60)外,所有参数几乎都具有最佳可靠性(ICC> 0.81)。除前房容积以及优越的虹膜角膜角和偏心率外,所有参数的重复性均极佳(COV <10%)。在最后三个参数中,重复性极限与平均值相比过高。结论健康年轻人中的GDSA在测量解剖参数,轴向参数,瞬时参数和屈光参数方面具有几乎完美的相关性,而周边术语的变异性更大。解剖参数(如测厚法,前房或虹膜角膜角度和偏心率)的可重复性受到限制。在健康的年轻人中,其他评估的参数具有很好的可重复性,并且它们的一致性限制显示了该设备的出色临床效果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号