首页> 外文学位 >Validation of Floc Size Measurements Made with Flow Particle Image Analyzer (FPIA)
【24h】

Validation of Floc Size Measurements Made with Flow Particle Image Analyzer (FPIA)

机译:使用流式颗粒图像分析仪(FPIA)进行的絮凝物尺寸测量的验证

获取原文
获取原文并翻译 | 示例

摘要

Clay flocs exist in different streams of the oil sands extraction process and in tailings treatment. The clay fraction in these streams has a significant effect on design and operation of slurry pipelines, flotation efficiency, sludging in the separation vessel, froth treatment and tailings behavior. The size of clay flocs plays an important role in determining the behavior of colloidal suspensions. The Flow Particle Image Analyzer (FPIA) is an image analysis based instrument used for particle size measurement. In this study, floc size measurements are made for suspensions of kaolinite in water. Size analysis results are additionally obtained using the Malvern Mastersizer and microscopy. The results of this study indicate that the Mastersizer is not an appropriate option for floc size measurement. However, the microscopy results were in good agreement with the FPIA measurements.
机译:粘土絮凝物存在于油砂提取过程的不同流和尾矿处理中。这些流中的粘土馏分对浆液管道的设计和操作,浮选效率,分离容器中的淤渣,泡沫处理和尾矿行为具有重大影响。粘土絮体的大小在确定胶体悬浮液的行为中起着重要作用。流式颗粒图像分析仪(FPIA)是基于图像分析的仪器,用于粒度测量。在这项研究中,对高岭石在水中的悬浮液进行了絮团尺寸测量。使用Malvern Mastersizer和显微镜还可以得到尺寸分析结果。这项研究的结果表明,Mastersizer不是絮凝物尺寸测量的合适选择。但是,显微镜结果与FPIA测量结果非常吻合。

著录项

  • 作者

    Saadat, Marzieh.;

  • 作者单位

    University of Alberta (Canada).;

  • 授予单位 University of Alberta (Canada).;
  • 学科 Chemical engineering.
  • 学位 M.S.
  • 年度 2014
  • 页码 239 p.
  • 总页数 239
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 老年病学;
  • 关键词

  • 入库时间 2022-08-17 11:40:47

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号