A wireless magnetoelastic glucose biosensor in blood plasma is described,based on using a mass sensitive magnetoelastic sensor as transducer.The glucose biosensor was fabricated by coating the ribbon-like,magnetoelastic sensor with a pH sensitive polymer and a biolayer of glucose oxidase(GOx)and catalase.The pH response polymer swells or shrinks,thereby changing sensor mass loading,respectively,in response to increase or decrease of pH values.The GOx-catalyzed oxidation of the glucose in blood plasma produces gluconic acid,resulting in the pH sensitive polymer shrinking,which in turn decreases the sensor mass loading.The results show that the proposed magnetoelastic glucose biosensor can be successfully applied to determine the concentration of glucose in blood plasma.At glucose concentration range of 2.5-20.0 mmol/l,the biosensor responses are reversible and linear,with a detection limit of 1.2 mmol/1.Since no physical connections between the sensor and the monitoring instruments are required,this proposed biosensor can potentially be applied to in vivo and in situ measurement of glucose concentration in physiological fluids.
机译:介绍了一种以质感磁弹性传感器为换能器的血浆中无线磁弹性葡萄糖生物传感器。该葡萄糖生物传感器是通过在带状磁弹性传感器上涂覆pH敏感聚合物和葡萄糖氧化酶(GOx)生物层制成的。 pH响应聚合物膨胀或收缩,从而响应于pH值的增加或降低而分别改变传感器的质量负载.GOx催化血浆中葡萄糖的氧化产生葡萄糖酸,导致pH敏感聚合物结果表明,所提出的磁弹性葡萄糖生物传感器可以成功地用于测定血浆中的葡萄糖浓度。在葡萄糖浓度范围为2.5-20.0 mmol / l时,生物传感器的响应为可逆且线性,检测极限为1.2 mmol / 1。由于传感器与监测仪器之间没有物理连接根据需要,该提议的生物传感器可以潜在地应用于体内和原位测量生理液中的葡萄糖浓度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号