首页> 外文期刊>Analytical chemistry >A PULSE AMPEROMETRIC SENSOR FOR THE MEASUREMENT OF ATMOSPHERIC HYDROGEN PEROXIDE
【24h】

A PULSE AMPEROMETRIC SENSOR FOR THE MEASUREMENT OF ATMOSPHERIC HYDROGEN PEROXIDE

机译:用于测量大气过氧化氢的脉冲安培传感器

获取原文
获取原文并翻译 | 示例
       

摘要

Gaseous H2O2 is sampled through a Nafion membrane diffusion scrubber while 1 mM HCl is maintained stationary in the scrubber. After a preselected preconcentration time (typically, 5-10 min), a valve is opened to allow the scrubber liquid to now by gravity over an electrochemical H2O2 sensor for a brief period. The miniature flow-over sensor consists of a Pt/Rh wire working electrode and a Pt wire counter electrode wound respectively on separate segments of a Nafion solid polymer electrolyte tubing supported on a Ag/AgCl wire reference electrode. A simple electronic interface and a personal computer are used to control and record the electrochemical measurement. The liquid phase detection limit for this sensor is similar to 30 nM H2O2 in the anodic oxidation mode. For a 9 min gas sample preconcentration period, the LOD (S/N = 3 criterion) is 0.11 ppbv H2O2(g). Ambient H2O2 data obtained with this instrument were in excellent agreement with those obtained by an established fluorometric technique in a blind intercomparison.
机译:通过Nafion膜扩散洗涤器采样气态H2O2,同时在洗涤器中保持1 mM HCl静止。在预选的预浓缩时间(通常为5-10分钟)后,打开阀门以使洗涤器液体在重力作用下在电化学H2O2传感器上短暂停留。微型溢流传感器由Pt / Rh焊丝工作电极和Pt焊丝反电极组成,分别缠绕在支撑在Ag / AgCl焊丝参比电极上的Nafion固体聚合物电解质管的不同段上。简单的电子界面和个人计算机用于控制和记录电化学测量值。在阳极氧化模式下,此传感器的液相检测极限类似于30 nM H2O2。在9分钟的气体样品预浓缩期间,LOD(S / N = 3标准)为0.11 ppbv H2O2(g)。用该仪器获得的环境H2O2数据与通过已建立的荧光技术在盲目比对中获得的数据非常吻合。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号