Lowering the detection limit of solvent polymeric ion-selective membrane electrodes. 2. Influence of composition of sample and internal electrolyte solution

机译:降低溶剂聚合物离子选择性膜电极的检测限。 2.样品和内部电解质溶液组成的影响

获取原文
获取原文并翻译 | 示例
       

摘要

The influence of the composition of the internal electrolyte solution on the response of Pb2+- and Ca2+-selective membrane electrodes is investigated. It is shown that the lower detection limit is improved by generating, in the membrane, ionic gradients that lead to a flux of primary ions toward the inner reference electrolyte solution. If the ion flux is too strong, it may cause analyte depletion at the membrane surface and, as a consequence, apparent super-Nernstian response. Such electrodes are not adequate to measure low analyte activities but can be used to determine unbiased selectivity factors. The results are interpreted in terms of a steady-state model, introduced in the companion paper, that describes the influence of concentration gradients generated by ion-exchange and coextraction processes on both sides of the membrane. [References: 20]
机译:研究了内部电解质溶液的组成对Pb2 +-和Ca2 +-选择性膜电极响应的影响。示出了通过在膜中产生离子梯度来提高检测下限,该离子梯度导致初级离子流向内部参比电解质溶液。如果离子通量过大,可能会导致膜表面的分析物耗尽,并因此导致明显的超能斯特响应。这种电极不足以测量低的分析物活性,但可用于确定无偏的选择性因子。结果用在随附论文中介绍的稳态模型来解释,该模型描述了离子交换和共萃取过程在膜的两侧产生的浓度梯度的影响。 [参考:20]