首页> 外文期刊>Electroanalysis >Steady-state model calculations predicting the influence of key parameters on the lower detection limit and ruggedness of solvent polymeric membrane ion-selective electrodes
【24h】

Steady-state model calculations predicting the influence of key parameters on the lower detection limit and ruggedness of solvent polymeric membrane ion-selective electrodes

机译:稳态模型计算,预测关键参数对溶剂聚合物膜离子选择电极的检测下限和坚固性的影响

获取原文
获取原文并翻译 | 示例
           

摘要

Using a previously described steady-state model, the influence of various parameters on the response function of ion-selective electrodes is investigated. The parameters studied include activities of interfering ions of the sample and inner solution, selectivity coefficients, lipophilicily of the analyte, concentrations of ionophore and lipophilic ionic sites in the membrane, the thickness of the Nernstian diffusion layer and the membrane as well as the diffusion coefficients in both phases. The results obtained provide indications as to how the lower detection limit of potentiometric sensors and their ruggedness can be improved. [References: 20]
机译:使用先前描述的稳态模型,研究了各种参数对离子选择电极响应函数的影响。研究的参数包括样品和内部溶液中干扰离子的活性,选择性系数,分析物的亲脂性,膜中离子载体和亲脂性离子位点的浓度,能斯登扩散层和膜的厚度以及扩散系数在两个阶段。获得的结果表明如何可以改善电位传感器的下检测限及其坚固性。 [参考:20]