Trace amounts of Hg2+ are detected by using a microcantilever coated with gold. The microcantilever undergoes bending due to accumulation of Hg2+ on the gold surface. It is found that a concentration of 10(-11) M Hg2+ can be detected using this technology. Other cations, such as K+, Na+ Pb2+, Zn2+, Ni2+, Cd2+, Cu2+, and Ca2+ have little or no effect on the deflection of the cantilever. The selectivity of the Hg2+ sensor could be improved by coating the gold surface of microcantilever with a self-assembled monolayer of a long-chain thiol compound. [References: 24]
机译:通过使用涂有金的微悬臂梁检测痕量的Hg2 +。由于Hg2 +在金表面的积累,微悬臂梁会发生弯曲。发现使用该技术可以检测到10(-11)M Hg2 +的浓度。其他阳离子,例如K +,Na +,Pb2 +,Zn2 +,Ni2 +,Cd2 +,Cu2 +和Ca2 +对悬臂的偏转几乎没有影响。通过用自组装的长链硫醇化合物单层覆盖微悬臂梁的金表面,可以提高Hg2 +传感器的选择性。 [参考:24]