Determination of oxide in fluoride salts using an yttria-stabilized-zirconia oxygen pump

机译:使用氧化钇稳定的氧化锆氧气泵测定氟盐中的氧化物

获取原文
获取原文并翻译 | 示例
       

摘要

An improved method over a previous technique has been developed to determine the ppm oxide concentration of fluoride salts. The oxide is released as oxygen gas by the reaction of the test salt with potassium bromotetrafluoride at 450 degrees C. The molecular oxygen released is then passed through a zirconia oxygen pump which selectively removes the oxygen. The current response is recorded as a chronoamperogram, from which the ppm oxide content of the salt can be obtained. Oxygen recovery from an yttrium oxide standard was better than 99%. The precision of analysis of FLINAK was better than 13% for samples containing 110-170 ppm oxide. The LOD was 12 mu g of oxygen, and the analytical range was 120 ppm to >20% for a 0.1-g sample. [References: 13]
机译:已经开发了一种相对于现有技术的改进方法来确定氟化物盐的ppm氧化物浓度。通过测试盐与溴化四氟化钾在450℃下的反应,氧化物以氧气的形式释放。释放出的分子氧然后通过氧化锆氧气泵,选择性地除去氧气。电流响应记录为计时电流图,从中可以得到盐的ppm氧化物含量。从氧化钇标准物中回收的氧气优于99%。对于含110-170 ppm氧化物的样品,FLINAK的分析精度优于13%。 LOD为12克氧气,0.1克样品的分析范围为120 ppm至> 20%。 [参考:13]