首页> 外文期刊>Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine >Development of epi-thermal neutron beam intensity detector with Ga-71(n,gamma)Ga-72 reaction for boron neutron capture therapy
【24h】

Development of epi-thermal neutron beam intensity detector with Ga-71(n,gamma)Ga-72 reaction for boron neutron capture therapy

机译:具有GA-71(N,γ)GA-72对硼中子捕获疗法的GA-71(N,GAMMA)GA-72反应的研制

获取原文
获取原文并翻译 | 示例

摘要

Based on the activation method using Ga-71(n,gamma)Ga-72 reaction, a cubic neutron flux intensity detector for epithermal neutrons was designed for boron neutron capture therapy (BNCT), and experimentally tested with a prototype detector in a neutron field produced at OKTAVIAN facility of Osaka University, Japan. The experimental test results and related analysis indicated that the performance of the detector was confirmed to be acceptable in the neutron field of BNCT. Practically, the neutron flux intensity mainly covering from 0.5 eV to 10 keV can be measured within 3% by the present detector.
机译:基于使用Ga-71(n,γ)Ga-72反应的激活方法,设计用于覆膜中子的立方中子磁通强度检测器,专为硼中子捕获治疗(BNCT),并用中子场中的原型检测器进行实验测试 在日本大阪大学的Oktavian设施中生产。 实验测试结果和相关分析表明,证实检测器的性能在BNCT的中子场中可接受。 实际上,主要覆盖0.5eV至10keV的中子磁通强度可以通过本探测器在3%以内测量。