An Ex Vivo Model in Human Femoral Heads for Histopathological Study and Resonance Frequency Analysis of Dental Implant Primary Stability

机译:人体股骨头的体内模型用于牙本质植入物原发稳定性的组织病理学研究和共振频率分析

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Objective. This study was designed to explore relationships of resonance frequency analysis (RFA)-assessed implant stability (ISQ values) with bone morphometric parameters and bone quality in an ex vivo model of dental implants placed in human femoral heads and to evaluate the usefulness of this model for dental implant studies. Material and Methods. This ex vivo study included femoral heads from 17 patients undergoing surgery for femoral neck fracture due to osteoporosis (OP) (n = 7) or for total prosthesis joint replacement due to severe hip osteoarthrosis (OA) (n = 10). Sixty 4.5 x 13 mm Dentsply Astra implants were placed, followed by RFA. CD44 immunohistochemical analysis for osteocytes was also carried out. Results. As expected, the analysis yielded significant effects of femoral head type (OA versus OA) (P < 0.001), but not of the implants (P = 0.455) or of the interaction of the two factors (P = 0.848). Bonferroni post hoc comparisons showed a lower mean ISQ for implants in decalcified (50.33 ± 2.92) heads than in fresh (66.93 ± 1.10) or fixated (70.77 ± 1.32) heads (both P < 0.001). The ISQ score (fresh) was significantly higher for those in OA (73.52 ± 1.92) versus OP (67.13 ± 1.09) heads. However, mixed linear analysis showed no significant association between ISQ scores and morphologic or histomorphometric results (P > 0.5 in all cases), and no significant differences in ISQ values were found as a function of the length or area of the cortical layer (both P > 0.08). Conclusion. Although RFA-determined ISQ values are not correlated with morphometric parameters, they can discriminate bone quality (OP versus OA). This ex vivo model is useful for dental implant studies.
机译:目的。这项研究旨在探讨共振频率分析(RFA)评估的植入物稳定性(ISQ值)与人股骨头植入牙的离体模型中骨形态参数和骨质量之间的关系,并评估该模型的实用性用于牙种植体研究。材料与方法。这项离体研究包括来自17名因骨质疏松症(OP)(n = 7)进行股骨颈骨折手术或因严重髋关节骨关节炎(OA)(n = 10)进行全修复关节置换术的患者的股骨头。放置60颗4.5 x 13毫米Dentsply Astra植入物,然后进行RFA。还对骨细胞进行了CD44免疫组化分析。结果。如预期的那样,该分析产生了股骨头类型的显着影响(OA与OA)(P <0.001),但对植入物(P = 0.455)或两个因素的相互作用(P = 0.848)却没有。 Bonferroni事后比较显示,脱钙(50.33±2.92)头的植入物的平均ISQ低于新鲜(66.93±1.10)或固定(7​​0.77±1.32)头的植入物(均P <0.001)。 OA组(73.52±1.92)的ISQ得分(新鲜)明显高于OP组(67.13±1.09)的ISQ得分(新鲜)。但是,混合线性分析显示,ISQ得分与形态或组织形态计量结果之间无显着关联(在所有情况下,P> 0.5),并且没有发现ISQ值与皮质层的长度或面积有显着差异(均P > 0.08)。结论。尽管RFA确定的ISQ值与形态参数无关,但它们可以区分骨骼质量(OP与OA)。该离体模型可用于牙种植体研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号