BACKGROUND: Several case reports have shown that the Pentax-AWS (Hoya Corporation, Tokyo, Japan), a new video laryngoscope, is useful in patients with difficult airways. METHODS: We assessed the effectiveness of the Pentax-AWS in two groups. Group 1 included 270 patients in whom direct laryngoscopy using a Macintosh laryngoscope had been difficult. Group 2 included 23 patients with predicted difficult intubation and difficult mask ventilation without previous use of the Macintosh laryngoscope. RESULTS: In group 1, the view of the glottis with the Macintosh laryngoscope was Cormack and Lehane grade 2 in 14 patients, grade 3 in 208 patients, and grade 4 in 48 patients. In 256 patients in whom the grade was 3 or 4 with the Macintosh laryngoscope, the view with the Pentax-AWS was either grade 1 or 2 in 255 patients (99.6%; 95% confidence intervals 97.8-100%). Tracheal intubation was successful with the Pentax-AWS in 268 of 270 patients (99.3%; 95% confidence interval 97.4-100%), and it failed (after two attempts) in two patients. In group 2, tracheal intubation was successful in 22 of 23 patients, and it failed in one patient. The reasons for failed intubation using the Pentax-AWS were failure to position the blade toward the glottic side of the epiglottis, inability to maneuver the endotracheal tube away from the arytenoids and into the trachea, and bleeding and swelling of the oropharynx. CONCLUSION: The success rate of tracheal intubation using the Pentax-AWS was high in patients with difficult laryngoscopy with a Macintosh laryngoscope and in patients with predicted difficult intubation.
机译:背景:一些病例报告表明,一种新型的视频喉镜“宾得-AWS”(Hoya Corporation,日本东京)可用于气道困难的患者。方法:我们评估了两组宾得AWS的有效性。第一组包括270例难以使用Macintosh喉镜进行直接喉镜检查的患者。第2组包括23名预计在不使用Macintosh喉镜的情况下插管困难和面罩通气困难的患者。结果:在第1组中,使用Macintosh喉镜检查的声门为14例患者的Cormack和Lehane 2级,208例患者为3级,48例患者为4级。在Macintosh喉镜中3级或4级的256位患者中,Pentax-AWS的视图在255位患者中为1级或2级(99.6%; 95%置信区间97.8-100%)。 Pentax-AWS在270例患者中的268例中成功进行了气管插管(99.3%; 95%的置信区间97.4-100%),而在2例患者中失败(两次尝试后)。在第2组中,气管插管在23例患者中的22例中成功,但在1例患者中失败。使用Pentax-AWS插管失败的原因有:无法将刀片朝会厌的声门侧定位;无法将气管导管从乙炔类药物转移到气管中;以及口咽部出血和肿胀。结论:在使用Macintosh喉镜的喉镜检查困难的患者和预计插管困难的患者中,使用Pentax-AWS进行气管插管的成功率很高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号