首页> 外文期刊>Biomedical Chromatography: An International Journal Devoted to Research in Chromatographic Methodologies and Their Applications in the Biosciences >Development and validation of an on-line two-dimensional reversed-phase liquid chromatography-tandem mass spectrometry method for the simultaneous determination of prostaglandins E(2) and F(2alpha) and 13,14-dihydro-15-keto prostaglandin F(2alpha) levels
【24h】

Development and validation of an on-line two-dimensional reversed-phase liquid chromatography-tandem mass spectrometry method for the simultaneous determination of prostaglandins E(2) and F(2alpha) and 13,14-dihydro-15-keto prostaglandin F(2alpha) levels

机译:在线二维反相液相色谱-串联质谱法同时测定前列腺素E(2)和F(2alpha)和13,14-dihydro-15-keto前列腺素F(2alpha)的开发和验证)水平

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

We developed and validated an on-line reverse-phase two-dimensional LC/MS/MS (2D-LC/MS/MS) system for simultaneous determination of the levels of prostaglandin (PG) E(2) as well as PGF(2alpha) and its metabolite 13,14-dihydro-15-keto PGF(2alpha) (F(2alpha)-M) in human plasma. Analytes were extracted by a three-step solid-phase extraction. Samples were then analyzed by on-line 2D-LC/MS/MS with electrospray ionization in negative mode. The 2D-LC system is composed of two reverse-phase analytical columns with a trapping column linking the two analytical columns. While an acidic buffer was used for both separation dimensions, differing organic solvents were employed for each dimension: methanol for the first and acetonitrile for the second to increase resolving power. The 2D-LC/MS/MS method was highly selective and sensitive with a significantly lower limit of quantitation (0.5 pg/mL for PGE(2) and 2.5 pg/mL for PGF(2alpha) and F(2alpha)-M, respectively). Linearity of the 2D-LC/MS/MS system was demonstrated for the calibration ranges of 0.5-50 pg/mL for PGE(2) and 2.5-500 pg/mL for PGF(2alpha) and F(2alpha)-M, respectively. Acceptable precision and accuracy were obtained throughout the calibration curve ranges. This highly selective and sensitive method was successfully utilized to determine the endogenous levels of PGE(2), PGF(2alpha), and F(2alpha)-M in plasma samples from six (four male and two female) normal volunteers. The mean concentrations for each analyte were 0.755 pg/mL for PGE(2), 5.70 pg/mL for PGF(2alpha) and 9.48 pg/mL for F(2alpha)-M.
机译:我们开发并验证了在线反相二维LC / MS / MS(2D-LC / MS / MS)系统,用于同时测定前列腺素(PG)E(2)和PGF(2alpha)的水平)及其在人体血浆中的代谢物13,14-dihydro-15-keto PGF(2alpha)(F(2alpha)-M)。通过三步固相萃取法萃取分析物。然后通过在线2D-LC / MS / MS在负模式下用电喷雾电离对样品进行分析。 2D-LC系统由两个反相分析柱和一个连接两个分析柱的捕集柱组成。虽然两个分离尺寸均使用酸性缓冲液,但每个尺寸均使用不同的有机溶剂:第一个用于分离甲醇,第二个用于分离乙腈,以提高分离能力。 2D-LC / MS / MS方法具有高度选择性和灵敏性,定量限较低(PGE(2)为0.5 pg / mL,PGF(2alpha)和F(2alpha-M)分别为2.5 pg / mL )。分别针对PGE(2)的0.5-50 pg / mL和PGF(2alpha)和F(2alpha-M)的2.5-500 pg / mL的校准范围证明了2D-LC / MS / MS系统的线性。在整个校准曲线范围内均获得了可接受的精度和准确性。这种高度选择性和敏感的方法已成功地用于确定六名(四名男性和两名女性)正常志愿者的血浆样品中PGE(2),PGF(2alpha)和F(2alpha)-M的内源水平。对于PGE(2),每种分析物的平均浓度为0.755 pg / mL,对于PGF(2alpha)为5.70 pg / mL,对于F(2alpha-M)为9.48 pg / mL。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号