Evaluation of the accuracy and reproducibility of a high-temperature differential scanning calorimeter by heat capacity measurements

机译:通过热容测量评估高温差示扫描量热仪的准确性和可重复性

获取原文
获取原文并翻译 | 示例
           

摘要

Experimental conditions were varied to study their effect on the accuracy and reproducibility of heat capacity measurements made with a high-temperature differential scanning calorimeter. Use of a higher heating rate, a calibration standard of similar size and mass of the sample, and a larger sample mass can give much more accurate and reproducible heat capacity measurements.
机译:改变实验条件以研究它们对用高温差示扫描量热仪进行的热容测量的准确性和可重复性的影响。使用更高的加热速率,具有相似尺寸和质量的样品的校准标准以及更大的样品质量可以提供更准确,可再现的热容测量。