A high-temperature adiabatic calorimeter has been developed to measure the constant-volume specific heat capacities (cV) of both gases and liquids, especially fluids of interest to emerging energy technologies. The chief design feature is its nearly identical twin bomb arrangement, which allows accurate measurement of energy differences without large corrections for energy losses due to thermal radiation fluxes. Operating conditions for the calorimeter cover a range of temperatures from 250 K to 700 K and at pressures up to 20 MPa. Performance tests were made with a sample of twice-distilled water. Heat capacities for water were measured from 300 K to 420 K at pressures to 20 MPa. The measured heat capacities differed from those calculated with an independently developed standard reference formulation with a root-mean-square fractional deviation of 0.48 %.
机译:已经开发出一种高温绝热量热仪来测量气体和液体,特别是新兴能源技术感兴趣的流体的恒定体积比热容(cV)。主要设计特征是其几乎相同的双炸弹布置,可以准确测量能量差,而无需对由于热辐射通量引起的能量损失进行大的校正。量热计的工作条件涵盖250 K至700 K的温度范围以及最高20 MPa的压力。使用两次蒸馏水的样品进行性能测试。在20 MPa的压力下,水的热容量从300 K到420K。测得的热容量不同于使用独立开发的标准参考配方计算的热容量,均方根分数偏差为0.48%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号