首页> 外文期刊>Journal of medical imaging and radiation sciences >Evaluation of the AAA Treatment Planning Algorithm for SBRT Lung Treatment: Comparison with Monte Carlo and Homogeneous Pencil Beam Dose Calculations
【24h】

Evaluation of the AAA Treatment Planning Algorithm for SBRT Lung Treatment: Comparison with Monte Carlo and Homogeneous Pencil Beam Dose Calculations

机译:SBRT肺部治疗AAA治疗规划算法的评价:与蒙特卡罗和均匀铅笔束剂量计算的比较

获取原文
获取原文并翻译 | 示例

摘要

Purpose: To evaluate the dose calculation accuracy of the Varian Eclipse anisotropic analytical algorithm (AAA) for stereotactic body radiation therapy (SBRT), and to investigate the dosimetric consequences of not applying tissue heterogeneity correction on complex SBRT lung plans.Materials and Methods: Nine cases of non-small-cell lung cancer (NSCLC) that were previously treated with SBRT at our center were selected for this study. Following Radiation Therapy Oncology Group 0236, the original plans were calculated using pencil beam without heterogeneity correction (PBNC) .For this study, these plans were recalculated by applying tissue heterogeneity correction with the AAA algorithm and with the Monte Carlo (MC) method, keeping the number of monitor units the same as the original plans. Two kinds of plan comparison were made. First, the AAA calculations were compared with MC.
机译:目的:评估瓦西亚蚀转视分析算法(AAA)对立体定向体放射治疗(SBRT)的剂量计算准确性,并研究了在复杂SBRT肺计划上不适用组织异质性校正的剂量后果。材料与方法:九 为本研究选择了先前用SBRT治疗的非小细胞肺癌(NSCLC)的病例。 在放射治疗肿瘤学组0236之后,使用铅笔梁计算出没有异质性校正(PBNC)的原始计划。对于本研究,通过使用AAA算法应用组织异质性校正和蒙特卡罗(MC)方法来重新计算这些计划,保持 监视器单元的数量与原始计划相同。 制作了两种计划比较。 首先,将AAA计算与MC进行比较。