首页> 外文期刊>Journal of medical imaging and radiation oncology >Assessment of voluntary deep inspiration breath‐hold with CINE CINE imaging for breast radiotherapy
【24h】

Assessment of voluntary deep inspiration breath‐hold with CINE CINE imaging for breast radiotherapy

机译:乳房放射治疗的Cine Cine成像评估自愿深吸气息呼吸

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Summary Deep Inspiration Breath‐Hold ( DIBH ) techniques for breast cancer radiation therapy ( RT ) have reduced cardiac dose compared to Free Breathing ( FB ). Recently, a voluntary deep inspiration breath‐hold ( vDIBH ) technique was established using in‐room lasers and skin tattoos to monitor breath‐hold. An in‐house quality assessment of positional reproducibility during RT delivery with vDIBH in patients with left‐sided breast cancer was evaluated. The electronic portal imaging device ( EPID ) was used in cinematographic ( CINE ) mode to capture a sequence of images during beam delivery. Weekly CINE images were retrospectively assessed for 20 left‐sided breast cancer patients receiving RT in vDIBH , and compared with CINE images of 20 patients treated in FB . The intra‐beam motion was assessed and the distance from the beam central axis ( CA ) to the internal chest wall ( ICW ) was measured on each CINE image. These were then compared to the planned distance on digitally reconstructed radiograph ( DRR ). The maximum intra‐beam motion for any one patient measurement was 0.30?cm for vDIBH and 0.20?cm for FB . The mean difference between the distance from the CA to ICW on DRR and the equivalent distance on CINE imaging (as treated) was 0.28?cm ( SD 0.17) for vDIBH patients and 0.25?cm ( SD 0.14) for FB patients ( P ?=?0.458). The measured values were comparable for patients undergoing RT in vDIBH , and for those in FB . This quality assessment showed that using in‐room lasers and skin tattoos to independently monitor breath‐hold in vDIBH as detected by ‘on‐treatment’ CINE imaging is safe and effective.
机译:发明内容乳腺癌放射治疗(RT)的深度灵感呼吸持有(DIBH)技术与自由呼吸(FB)相比,心脏剂量降低了。最近,使用室内激光器和皮肤纹身来确定一个自愿的深度灵感呼吸阻滞(VDIBH)技术,以监测呼吸持有。评价左侧乳腺癌患者RT递送在RT递送期间存在的内部质量评估。电子门户成像装置(EPID)用于电影摄影(CINE)模式以在光束输送期间捕获一系列图像。回顾性地评估每周的左侧乳腺癌患者在vdibh中的左侧乳腺癌患者,并与FB治疗的20名患者的调解图像进行比较。在每个Cine图像上测量横梁内运动,并在每个Cine图像上测量与梁中心轴(CA)到内胸壁(ICW)的距离。然后将这些与在数字重建的射线照片(DRR)上的计划距离进行比较。对于任何一个患者测量的最大束射线运动对于VDIBH和FB的0.20Ωcm为0.30Ωcm。对于VDIBH患者的VDIBH患者,从DRR上的CA至ICW与ICW的距离与CARR上的距离和相同距离(如处理)为0.28Ωcm(SD 0.17),FB患者为0.25Ωcm(SD 0.14)(P?= ?0.458)。测量值对于在VDIBH中进行RT的患者以及FB中的患者进行了比较。这种质量评估表明,使用在“治疗”的CINE成像中检测到的VDIBH中的室内激光和皮肤纹身在VDIBH中独立监测呼吸率,这是安全有效的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号