In Vitro Performance Testing of Legacy and ENFit Gastrostomy Tube Devices Under Gravity Flow Conditions

机译:在重力流动条件下的遗留和基因胃术管装置的体外性能测试

获取原文
获取原文并翻译 | 示例
           

摘要

Background: Changes in connector design for the gastrostomy tube were implemented to reduce the risk of misconnections. This study aimed to determine whether there are differences in gravity flow rates between legacy devices and the ENFit devices intended to replace them. Materials and Methods: We compared 5 legacy gastrostomy tube brands with 3 corresponding ENFit brands, sized between 14 French (Fr) and 24 Fr. Seven commercial diets were used. One comparison involved low-profile devices. Results: Whether an ENFit device manifested a lower flow rate than a legacy device was not a strong function of diet. One 14-Fr ENFit device, because of its reduced distal inner tube diameter, produced an average feeding time of 56 (±13) minutes from a 20-minute baseline. For other 14-Fr ENFit devices, the increase was much less pronounced (25 ± 4 minutes). At larger sizes, both decreases and increases in feeding time were observed, depending on device type; on average, the 20-minute feeding time increased to 25 (±7) minutes. For low-profile devices, across all sizes, an increase in 20-minute feeding time occurred, but the difference was small (23 ± 2). Conclusion: Statistically lower flow rates were observed for 70% of ENFit devices relative to their legacy counterparts. We estimate that 30% of the differences may be noticeable. In the scenarios studied, lower flow rates (relative to other devices at the same Fr number) arise from energy losses in straight tubing. This difference can be reduced by increasing the tube inner diameters in distal end of ENFit tubes.
机译:背景:实施胃造口管的连接器设计的变化,以降低误差的风险。本研究旨在确定遗留设备和旨在更换它们的绑定装置之间是否存在重力流速的差异。材料和方法:我们将5种遗留胃术管品牌进行比较,其中3个相应的依托品牌,14个法国(FR)和24 Fr之间的大小。使用了七种商业饮食。一个比较涉及低调设备。结果:是否表现出比传统设备的流量较低的流量,而不是饮食的强效果。一个14-FR烯烯件装置,因为其远端管直径减小,从20分钟的基线产生56(±13)分钟的平均喂食时间。对于其他14-FR绑定器件,增加的速度远低得多(25±4分钟)。在较大的尺寸下,根据器件类型观察到喂食时间的降低和增加;平均而言,20分钟的喂食时间增加到25(±7)分钟。对于低调设备,在所有尺寸上,发生20分钟的喂食时间,但差异很小(23±2)。结论:统计上较低的流量率为70%相对于其遗留对应物。我们估计了30%的差异可能是显着的。在所研究的情况下,从直线管中的能量损耗发生较低的流速(相对于相同FR编号的其他器件)。通过增加依烯件管的远端管内径来降低这种差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号