首页> 外文期刊>Journal of the American Pharmacists Association: JAPhA >Clinical effects of calibration of CoaguChek XS PT Test Strips on Anticoagulation Management Services
【24h】

Clinical effects of calibration of CoaguChek XS PT Test Strips on Anticoagulation Management Services

机译:Coaguchek XS PT试验条对抗凝管理服务的临床疗效

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Objective: In January 2018, Roche Diagnostics calibrated CoaguChek XS PT Test Strips to the latest World Health Organization reference preparation. On September 12, 2018, Roche Diagnostics issued an urgent medical device correction statement warning of inaccurate international normalized ratio (INR) results greater than 4.5 and recommended additional confirmatory testing. Eight weeks later, on November 2, the U.S. Food and Drug Administration issued a class I recall of all affected lot numbers. The objective of this project was to determine the clinical effects of recalibration of CoaguChek on patients managed by the Anticoagulation Management Services (AMS).
机译:目的:2018年1月,Roche Diagnostics校准了Coaguchek XS PT测试条,以最新的世界卫生组织参考准备。 2018年9月12日,Roche Diagnostics发出了紧急医疗器械矫正声明,警告不准确的国际标准化比率(INR)结果大于4.5,推荐的额外确认测试。 八周后,11月2日,美国食品和药物管理局发布了一级,我回忆起所有受影响的批号。 该项目的目的是确定Coaguchek对由抗凝管理服务(AMS)管理的患者的临床影响。