European Concerted Action on Anticoagulation (ECAA). An assessment of lyophilised plasmas for ISI calibration of CoaguChek and TAS whole blood prothrombin time monitors

机译:欧洲抗凝协调行动(ECAA)。对冻干血浆的评估用于对CoaguChek和TAS全血凝血酶原时间监测仪进行ISI校准

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Aims: The recommended method for the international sensitivity index (ISI) calibration of whole blood point of care testing (POCT) prothrombin time (PT) systems was originally described by Tripodi et al in 1993 but is too complex and demanding. The present European Concerted Action on Anticoagulation (ECAA) study aimed to assess the reliability of simpler ISI calibration using lyophilised plasma samples.>Methods: ISI calibrations using three different types of ECAA lyophilised plasma samples (artificially depleted, individual, and pooled coumarin) were compared with whole blood calibrations on CoaguChek Mini and TAS PT-NC POCT monitors at 10 centres.>Results: With CoaguChek Mini systems, lyophilised coumarin plasma samples (both single donation and pooled) gave ISI and international normalised ratio (INR) values comparable to whole blood. With artificially depleted plasma, ISI and INR values were too high. With TAS PT-NC systems, all three types of lyophilised plasma samples gave inaccurate ISI and unreliable INR results, similar to previous ECAA findings with fresh plasma calibrations.>Conclusions: With CoaguChek Mini systems, ISI calibration can be simplified by the use of ECAA lyophilised plasma samples from coumarin treated patients. Further study is needed to devise a simpler calibration method for the TAS PT-NC system.
机译:>目的:最初由Tripodi等人在1993年描述了推荐的全血即时检测点(POCT)凝血酶原时间(PT)系统国际敏感性指数(ISI)校准的方法,但该方法过于复杂和苛刻。当前的欧洲抗凝协调行动(ECAA)研究旨在评估使用冻干血浆样品进行简单ISI校准的可靠性。>方法:使用三种不同类型的ECAA冻干血浆样品(人工消耗,单独使用)进行ISI校准和合并的香豆素)与CoaguChek Mini和TAS PT-NC POCT监测仪在10个中心的全血校准值进行了比较。>结果:使用CoaguChek Mini系统,冻干的香豆素血浆样品(单次捐赠和合并)得出的ISI和国际标准化比率(INR)值可与全血媲美。对于人工消耗的血浆,ISI和INR值太高。使用TAS PT-NC系统时,所有三种类型的冻干血浆样品均给出了不准确的ISI和不可靠的INR结果,这与以前使用新鲜血浆校准法得到的ECAA结果相似。通过使用经香豆素治疗的患者的ECAA冻干血浆样品简化了操作。需要为TAS PT-NC系统设计一种更简单的校准方法的研究。

著录项

相似文献

  • 外文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号