Radiolucency in stemless shoulder arthroplasty is associated with an imaging phenomenon

机译:在尾肩关节术中的辐射性与成像现象有关

获取原文
获取原文并翻译 | 示例
           

摘要

ABSTRACT Stemless humeral implants show comparable midterm clinical results compared to stemmed components. Recently, radiolucencies around the metaphyseal seating of humeral stemless implants were reported on postoperative radiographs. It is controversial whether they are attributable to bone resorption. We hypothesized these radiolucencies result from imaging artifacts. Seven cadaveric specimens (three male and four female) were first radiographed and then scanned with CT. A stemless humeral component of current design was implanted in each specimen. After implantation, all specimens were radiographed with different exposure settings. The implant was removed, and the specimens were scanned with CT again. Pre‐ and post‐implantation radiographs and CT scans were compared. The mean Hounsfield units (HU) at the humeral resection plane from the pre‐implantation CT were correlated with the diameter of the radiolucent halo on the post‐implantation radiographs. A symmetric radiolucent halo of variable diameters occurred on all radiographs after implantation when an automatic exposure control was used. The halo disappeared in all specimens when the tube voltage was reduced. Lower CT‐values (HU) before the implantation resulted in greater halos on the radiograph after implantation. Symmetric radiolucent halos can result from imaging artifacts, which is most likely due to radiation scatter. The halos can be minimized by reducing the tube voltage. The halo effect appears to be pronounced in bones with decreased density. ? 2017 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 35:2040–2050, 2017.
机译:摘要无尾肱骨植入物显示与茎组件相比的相当的中期临床结果。最近,在术后射线照片上报道了肱骨疏松植入物周围的辐射术。无论它们是否归因于骨吸收,它都是争议的。我们假设这些瑞利可造成的成像伪影。七个尸体标本(三个雄性和四个雌性)首次射线摄影,然后用CT扫描。在每个样品中植入电流设计的无尾肱骨部件。植入后,所有标本都以不同的曝光设置射线照相。除去植入物,再次用CT扫描样品。比较预植入后和植入后的X线检查和CT扫描。从预注入的PT的肱骨切除平面上的平均Hounsfield单元(Hu)与植入后射线照片上的无线卤素的直径相关。当使用自动曝光控制后,在植入后的所有射线照片上发生了对称的无线卤素的晕晕。当管电压降低时,光环在所有样本中消失。在植入后植入前降低CT值(HU),在植入后射线照片导致更大的卤素。对称的无辐射晕晕可以是由成像伪影产生的,这很可能是由于辐射散射而导致的。通过降低管电压,可以最小化晕圈。晕圈效应似乎在骨骼中发音,密度降低。还2017年骨科研究会。由Wiley Hearyicals,Inc.J orthop Res 35:2040-2050,2017出版。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号