Summary F?rster resonance energy transfer (FRET) probes being used to improve the resolution of stimulated emission depletion (STED) microscopy are numerically discussed. Besides the FRET efficiency and the excitation intensity, the fluorescence lifetimes of donor and acceptor are found to be another key parameter for the resolution enhancement. Using samples of FRET pairs with shorter donor lifetime and longer acceptor lifetime enhances the nonlinearity of the donor fluorescence, which leads to an increased resolution. The numerical simulation shows that a double resolution improvement of STED microscopy can be achieved by using Cy3–Atto647N samples when compared with that of using standard Cy3‐only samples.
机译:概述F?用于改善刺激发射耗尽(STED)显微镜的分辨率用于改善刺激的发射耗尽(STED)显微镜的分辨率的F?数值讨论。 除了FRET效率和励磁强度之外,还发现供体和受体的荧光寿命是分辨率增强的另一个关键参数。 使用具有较短供体寿命和更长的受体寿命的褶皱对的样品增强了供体荧光的非线性,这导致了增加的分辨率。 数值模拟表明,与使用标准CY3的样品相比,通过使用Cy3-Atto647N样品,可以通过使用Cy3-Atto647N样品来实现分层显微镜的双分辨率改善。