首页> 外文期刊>Journal of Low Temperature Physics >LC Filters for FDM Readout of the X-IFU TES Calorimeter Instrument on Athena
【24h】

LC Filters for FDM Readout of the X-IFU TES Calorimeter Instrument on Athena

机译:用于FDM读数的LC滤波器在Athena上的X-IFU TES卡路仪仪器

获取原文
获取原文并翻译 | 示例
           

摘要

The current status of lithographic superconducting LC filters for use in the Athena-X-IFU instrument is described. We present the fabrication process and characterization results at room temperature, 4?K and 50?mK. We also present an optimization study of the quality topics, where finite element modeling is used together with experimental validation structures. For the a-Si:H-based capacitors and Nb-based coils, presently the component fabrication yield is about 99% and the effective series resistance at?50 mK is lower than 1.5?mΩ.
机译:描述了用于Athena-X-IFU仪器的光刻超导LC过滤器的当前状态。 我们介绍了在室温下的制造过程和表征结果,4?K和50?MK。 我们还提供了对优质主题的优化研究,其中有限元建模与实验验证结构一起使用。 对于A-Si:基于H基电容器和基于Nb的线圈,目前部件制造产率约为99%,并且在Δ50mk处的有效串联电阻低于1.5ΩmΩ。