Absolute Steady-State Thermal Conductivity Measurement of insulation Materials Using Square Guarded Hot Plate Apparatus

机译:绝缘稳态热导率测量绝缘材料使用方形防护热板装置

获取原文
获取原文并翻译 | 示例
           

摘要

An absolute steady-state thermal conductivity measurement of fibrous and non-fibrous insulation materials is carried out using Square Guarded Hot Plate (SGHP) apparatus. The light weight fibrous insulating blankets (25 kg/m{sup}3 to 190 kg/m{sup}3) as well as the heavy insulating slabs (1800 kg/m{sup}3 to 2000 kg/m{sup}3) of 300 mm × 300 mm size specimens are considered for the thermal conductivity measurement. The thermal conductivities of fibrous insulation blanket and heavy insulating plates are found in the range of 0.0357 to 0.112 ± 0 0028 W/m.K and 0.050 to 0.239 ± 0.0048 W/m.K respectively. The thermal conductivity data obtained from measurements are compared with analytical model for various heavy insulating plates. The results are in good agreement with the analytical model with maximum deviation of ± 21 % for heavy insulating plates. An uncertainty analysis yields a combined measurement uncertainty of ± 5.75 % at no flow condition. The steady-state square guarded hot plate apparatus can be used effectively to estimate the effective thermal conductivity for all types of insulation materials.
机译:使用Square Lugened Hot Plate(SGHP)装置进行纤维和非纤维绝缘材料的绝对稳态导热性测量。轻质纤维绝缘橡皮布(25kg / m {sup} 3至190kg / m {sup} 3)以及重绝缘板(1800kg / m {sup} 3至2000kg / m {sup} 3) )300mm×300mm尺寸的标本被认为是导热性测量的。纤维绝缘橡皮布和重绝缘板的导热率在0.0357至0.112±0 0028W / m.K的范围内,分别为0.050至0.239±0.0048W / m.K。与各种重绝缘板的分析模型进行比较从测量中获得的导热系数。结果与分析模型吻合良好,其重型绝缘板的最大偏差为±21%。不确定性分析在没有流动条件下产生±5.75%的组合测量不确定度。稳态方形保护的热板装置可以有效地使用,以估计所有类型的绝缘材料的有效导热率。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号