首页> 外文期刊>Journal of applied microbiology >BIOLOGICAL INDICATORS FOR LOW TEMPERATURE STEAM AND FORMALDEHYDE STERILIZATION - EFFECT OF VARIATIONS IN RECOVERY CONDITIONS AN THE RESPONSE OF SPORES OF BACILLUS STEAROTHERMOPHILUS NCIMB 8224 TO LOW TEMPERATURE STEAM AND FORMALDEHYDE
【24h】

BIOLOGICAL INDICATORS FOR LOW TEMPERATURE STEAM AND FORMALDEHYDE STERILIZATION - EFFECT OF VARIATIONS IN RECOVERY CONDITIONS AN THE RESPONSE OF SPORES OF BACILLUS STEAROTHERMOPHILUS NCIMB 8224 TO LOW TEMPERATURE STEAM AND FORMALDEHYDE

机译:低温蒸汽和甲醛灭菌的生物指标 - 恢复条件变化的影响芽孢杆菌孢子菌孢子8224至低温蒸汽和甲醛的响应

获取原文
获取原文并翻译 | 示例
           

摘要

Spores of a Bacillus stearothermophilus strain thought to be a good biological monitor for low temperature steam and formaldehyde (LTSF) sterilization have been subjected to different recovery environments following exposure to relevant LTSF treatments and prior to enumerating the numbers of surviving spores. The recovery treatments essentially consisted of holding samples at a temperature of 90 degrees C for different time periods prior to plating out following exposure to 12 mu g ml(-1) formaldehyde at temperatures between 63 degrees and 83 degrees C. The data indicate that LTSF-exposed spores show a marked recovery when kept at 90 degrees C prior to plating out; the extent of the recovery increases with increasing inactivation temperature between 73 degrees and 83 degrees C. The data bring into question the sporicidal activity of LTSF.
机译:芽孢杆菌的孢子菌株认为是低温蒸汽和甲醛(LTSF)灭菌的良好的生物监测液,在接触相关的LTSF治疗后经受不同的回收环境,并在列举幸存孢子的数量之前进行不同的回收环境。 恢复处理基本上由在在暴露于12μg(-1)甲醛之前在63度和83摄氏度之间的温度下电镀后的不同时间段在90℃的温度下保持样品。数据表明LTSF 佩齐的孢子显示在电镀之前保持在90度的标记恢复; 回收的程度随着73度和83摄氏度之间的不变温度而增加而增加。数据归因于质疑LTSF的Sporicatal活性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号