首页> 外文期刊>Journal of cardiovascular translational research >Evaluation of the CorInnova Heart Assist Device in an Acute Heart Failure Model
【24h】

Evaluation of the CorInnova Heart Assist Device in an Acute Heart Failure Model

机译:急性心脏衰竭模型中Corinnova心脏辅助装置的评价

获取原文
获取原文并翻译 | 示例
       

摘要

While the number of patients supported with temporary cardiac assist is growing, the existing devices are limited by a multitude of complications, mostly related to contact with the blood. The CorInnova epicardial compressive heart assist device was tested in six sheep using an acute heart failure model. High esmolol dose, targeting a 50% reduction in CO from healthy baseline, resulted in a failure state with mean CO 1.9L/min. Heart assist with the device during failure state resulted in an average absolute increase in CO of 1.0L/min, along with a decline in ventricular work to 67.5% of the total LV SW. Combined with repeated success of minimally invasive device implant, the resulting increases in cardiac hemodynamics achieved while still unloading the heart demonstrate the potential of the CorInnova device for temporary heart assist.
机译:虽然临时心脏辅助支持的患者数量不断增长,但现有设备受到多重并发症的限制,主要是与血液接触有关。 使用急性心力衰竭模型在六只绵羊中测试了Corinnova心外膜压缩心脏辅助装置。 高ESMOLOL剂量,靶向来自健康基线的CO的50%,导致具有平均CO 1.9L / min的失效状态。 心脏辅助装置在失效状态下的辅助导致1.0L / min的CO的平均绝对增加,并且心室工作下降至总LV SW的67.5%。 结合微创装置植入物的重复成功,所得心脏血流动力学的产生增加,同时仍然卸载心脏展示了Corinnova装置的临时心脏辅助的潜力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号