Validation of a noninvasive blood pressure monitoring device in normotensive and hypertensive pediatric intensive care patients.

机译:规范性高血压儿科重症监护患者的非血压血压监测装置的验证。

获取原文
获取原文并翻译 | 示例
       

摘要

OBJECTIVE: To evaluate the performance and to define limitations of a noninvasive blood pressure monitoring device in the critically ill pediatric population. METHOD: Patients were included in the study if they were admitted to the Pediatric Intensive Care Unit, were between the ages of 1 month and 18 years with wrist circumferences of > or =10 cm, and had an indwelling arterial line. Patients were excluded if their systolic blood pressure differed by > or =7.5% between their upper extremities. The measurements were collected simultaneously with those from an arterial line by a computer interfaced with the noninvasive blood pressure monitoring system and the patient's monitor. Heart rates were calculated from the recorded pulse waveforms of the arterial lines. Comparison analyses were performed via bias and precision plots of the blood pressure and heart rate data in addition to calculation of Pearson's correlation coefficients and concordance correlation coefficients. As a nonparametric method of comparison, the proportion of measurements that differed by greater than 10% was calculated. Results. Blood pressures and heart rates of 20 patients between the ages of 12 months and 17 years were monitored by a noninvasive blood pressure monitor for 30 min per patient. This data collection resulted in 2015 data points for each blood pressure and heart rate for comparison of methods. Concordance correlation coefficients were the following: systolic blood pressure, 0.93; diastolic blood pressure, 0.93; mean blood pressure, 0.94; and heart rate, 0.85. CONCLUSIONS: The noninvasive blood pressure monitor is capable of producing an accurate blood pressure measurement every 12-15 heartbeats in addition to providing a pulse waveform and digital display of the heart rate. Our study showed good agreement between the methods in the normotensive and hypertensive critically ill pediatric population with a wrist circumference limitation defined at > or =11 cm.
机译:目的:评价性能和定义批评性儿科人群中的非血压血压监测装置的局限性。方法:患者纳入研究中,如果他们被录取到儿科重症监护病房,在1个月和18岁之间的腕骨周长均为>或= 10厘米,并具有留置动脉线。如果它们的上肢之间的收缩压差异不同,则排除患者。通过与非侵入性血压监测系统和患者显示器接口的计算机将测量与来自动脉线的电脑同时收集。从动脉线的记录脉冲波形计算心率。除了计算Pearson的相关系数和一致性相关系数之外,还通过血压和心率数据的偏置和精密图进行比较分析。作为非参数的比较方法,计算了不同大于10%的测量比例。结果。每位患者通过非血压血压监测器监测12个月和17岁的20岁患者的血压和心率。该数据收集导致2015年为每种血压和心率进行的数据点进行比较。一致性相关系数如下:收缩压,0.93;舒张压,0.93;平均血压,0.94;和心率,0.85。结论:除了提供心率的脉冲波形和数字显示器之外,非侵入性血压监测器还能够每12-15心跳产生精确的血压测量。我们的研究表明,通过>或= 11厘米的腕圆周限制,患有腕圆周的儿科人群的正常和高血压批评的方法之间的研究表明。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号