Limitations of lesion quality estimated by ablation index: An in vitro study

机译:消融指标估计的病变质量的限制:体外研究

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Limitations of the ablation index background Ablation index (AI) is a novel marker of lesion quality from radiofrequency (RF) catheter ablation. However, AI reliability has not been fully validated by experimental data. The aim of the present study is to validate AI reliability for estimating lesion size using different settings for RF parameters: contact angle, power delivery, and contact force (CF). Methods and results We evaluated the lesion size in porcine hearts (N = 108) after RF application at three different contact angles to the myocardium: perpendicular (90 degrees), oblique (45 degrees), and parallel (0 degrees). At each angle, RF power at 25, 30, and 35 W was applied at target CF values of 5, 15, and 30 g as measured by the CF sensor to reach target AIs of 300, 400, 500, and 600. AI value was significantly correlated with lesion depth, width, and volume (R = 0.84, 0.82, and 0.87, respectively, all P < 0.001). Lesion depth decreased with smaller contact angles (45 degrees and 0 degrees). Furthermore, high-power RF energy (35 W) resulted in a significantly smaller lesion volume compared with standard-power energy (30 W). There were no significant differences in lesion size among CF settings. Conclusions AI was strongly correlated with lesion depth, width, and volume, but only within a small range of contact angles and RF power delivery settings.
机译:消融指标背景消融指标(AI)的局限性是来自射频(RF)导管消融的损伤质量的新标记。但是,通过实验数据尚未完全验证AI可靠性。本研究的目的是使用不同的RF参数估算病变尺寸来验证AI可靠性:接触角,电力输送和接触力(CF)。方法和结果我们在RF施加在三个不同的接触角度到心肌的猪心中(n = 108)中的病变大小评价:垂直(90度),斜(45度)和平行(0度)。在每个角度下,通过CF传感器测量的5,15和30g的靶CF值,施用25,30和35W的RF功率,以达到300,400,500和600的目标AIS。ai值与病变深度,宽度和体积显着相关(r = 0.84,0.82,0.87,所有P <0.001)。损伤深度随着较小的接触角(45度和0度)而降低。此外,与标准功率能量(30W)相比,高功率RF能量(35W)导致显着更小的病变量。 CF设置中的病变大小没有显着差异。结论ai与病变深度,宽度和体积强烈相关,但仅在少量接触角和射频电力输送环境范围内。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号