首页> 外文期刊>Journal of Biomechanics >Radiostereometric analysis using clinical radiographic views: Development of a universal calibration object
【24h】

Radiostereometric analysis using clinical radiographic views: Development of a universal calibration object

机译:使用临床射线照相观点的radioRTere000测量分析:普遍校准对象的开发

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Radiostereometric analysis (RSA) is a highly accurate technique used to provide three-dimensional (3D) measurements of orthopaedic implant migration for clinical research applications, yet its implementation in routine clinical examinations has been limited. Previous studies have introduced a modified RSA procedure that separates the calibration examinations from the patient examinations, allowing routine clinical radiographs to be analyzed using RSA. However, in order to calibrate the wide range of clinical views, a new calibration object is required. In this study, a universal, isotropic calibration object was designed to calibrate any pair of radiographic views used in the clinic for RSA. A numerical simulation technique was used to design the calibration object, followed by a phantom validation test of a prototype to verify the performance of the novel object, and to compare the measurement reliability to the conventional calibration cage. The 3D bias for the modified calibration method using the new calibration object was 0.032 +/- 0.006 mm, the 3D repeatability standard deviation was 0.015 mm, and the 3D repeatability limit was 0.042 mm. Although statistical differences were present between the universal calibration object and the conventional cage, the differences were considered to be not clinically meaningful. The 3D bias and repeatability values obtained using the universal calibration object were well under the threshold acceptable for RSA, therefore it was successfully validated. The universal calibration object will help further the adoption of RSA into a more routine practice, providing the opportunity to generate quantitative databases on joint replacement performance. (C) 2018 Elsevier Ltd. All rights reserved.
机译:辐射静脉分析(RSA)是一种高精度的技术,用于提供用于临床研究应用的矫形植入物迁移的三维(3D)测量,但其在常规临床检查中的实施得到了限制。以前的研究引入了一种修饰的RSA程序,将校准检查与患者检查中的校准检查分开,允许使用RSA分析常规的临床射线照片。但是,为了校准各种临床视图,需要一种新的校准对象。在本研究中,旨在校准通用的各向同性校准对象以校准诊所用于RSA的任何对射线照相视图。使用数值模拟技术来设计校准对象,然后是原型的幻影验证测试,以验证新型对象的性能,并将测量可靠性与传统校准笼进行比较。使用新校准对象的改进校准方法的3D偏置为0.032 +/- 0.006 mm,3D可重复性标准偏差为0.015毫米,3D可重复限制为0.042毫米。虽然普遍校准对象和传统笼之间存在统计差异,但差异被认为是在临床上有意义的。使用通用校准物体获得的3D偏差和可重复性值在RSA可接受的阈值下井井状线,因此它已成功验证。通用校准对象将进一步通过RSA进入更多例程实践,提供有机会在联合置换性能上产生定量数据库。 (c)2018年elestvier有限公司保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号