Calibration of Stereo Radiography System for Radiostereometric Analysis Application

机译:立体放射成像系统在放射立体分析中的应用校准

获取原文

摘要

This paper describes a new alternative to conventional radiography system currently used for radiostereometric analysis studies. Instead of using two non-calibrated X-ray sources with a cumbersome calibration cage, we propose to use the biplanar radiography EOS system. Its fixed configuration provides a preliminary calibration and a much simpler acquisition protocol. A flexible and accurate calibration method is presented to optimize EOS default calibration using a simple object and a self-calibration method. To validate our system, we calculate the 3D reconstruction error of a known object. Results showed an accuracy of 70±11μm and 0.05±0.02° for translation and rotation respectively, and an average epipolar error of 23±03μm.
机译:本文介绍了目前用于放射线立体分析研究的常规放射线照相系统的新替代品。我们建议使用双平面X射线照相EOS系统,而不是使用两个笨重的校准笼来校准X射线源。它的固定配置提供了初步的校准和更简单的采集协议。提出了一种灵活而准确的校准方法,以使用简单的对象和自校准方法来优化EOS默认校准。为了验证我们的系统,我们计算了已知物体的3D重建误差。结果表明,平移和旋转的精度分别为70±11μm和0.05±0.02°,平均对极误差为23±03μm。