首页> 外文期刊>Journal of AOAC International >Validation of the BetaStar (R) Advanced for Tetracyclines Test Kit for the Screening of Bulk Tank and Tanker Truck Milks for the Presence of Tetracycline Drug Residues
【24h】

Validation of the BetaStar (R) Advanced for Tetracyclines Test Kit for the Screening of Bulk Tank and Tanker Truck Milks for the Presence of Tetracycline Drug Residues

机译:用于筛选四环素罐和油轮卡车阵雨的四环素试剂盒的效果,用于存在四环素药物残留物

获取原文
获取原文并翻译 | 示例
           

摘要

A validation study was conducted for an immunochromato-graphic method (BetaStar (R) Advanced for Tetracyclines) for detection of tetracycline antibiotic residues in raw, commingled bovine milk. The assay was demonstrated to detect tetracycline, chlortetracycline, and oxytetracycline at levels below the FDA tolerance levels but above the maximum sensitivity thresholds established by the National Conference on Interstate Milk Shipments. Results of internal and independent laboratory dose response studies employing spiked samples were in agreement. All three drugs at the approximate 90/95% sensitivity levels were detected in milk collected from cows that had been treated with the specific drug. Selectivity of the assay was 100%, as no false-positive results were obtained in testing 881 control milk samples. Testing the estimated 90/95 sensitivity level for tetracycline (213 ppb), chlortetracycline (272 ppb), and oxytetracycline (180 ppb) and at 1000 ppb for each antibiotic resulted in 100% positive tests for each tetracycline. Results of ruggedness experiments established the operating parameter tolerances for the test. Results of cross-reactivity testing established that the assay detects certain other tetracycline drugs but does not cross-react with any of 32 drugs belonging to seven different drug classes. Abnormally high bacterial or somatic cell counts (SCC) in raw milk produced no assay interference.
机译:进行验证研究以进行免疫噻唑 - 图形方法(用于四环素的甜菜碱(R)),用于检测原料,混合牛奶中的四环素抗生素残留物。证明该测定以检测四环素,氯化萘曲线和催眠素,低于FDA耐受水平,但高于全国州际奶制股票会议建立的最大敏感阈值。使用尖刺样品的内部和独立实验室剂量反应研究的结果达成一致。在由特定药物处理的奶牛收集的牛奶中检测到近似90/95%敏感性水平的所有三种药物。测定的选择性为100%,因为在测试881对照牛奶样品中没有获得假阳性结果。测试四环素(213ppb),氯化碳酸碱(272ppb)和氧赤素(180ppb)和每种抗生素的1000ppb的估计的90/95个灵敏度水平导致每种四环素的100%阳性试验。坚固耐力实验结果建立了测试的操作参数公差。交叉反应性试验的结果确定了测定检测到某些其他四环素药物,但不会与属于七种不同药物课程的32种药物中的任何一种交叉反应。原料乳中的异常高细菌或体细胞计数(SCC)没有产生测定干扰。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号