首页> 外文期刊>Diagnostic microbiology and infectious disease >Automated broth-based systems versus the MYCOTB plate for antimicrobial susceptibility testing of the Mycobacterium tuberculosis complex: challenges in interpretation
【24h】

Automated broth-based systems versus the MYCOTB plate for antimicrobial susceptibility testing of the Mycobacterium tuberculosis complex: challenges in interpretation

机译:基于自动化的肉汤系统与Mycotb板进行抗菌易感性测试的分枝杆菌复合体复合体:解释中的挑战

获取原文
获取原文并翻译 | 示例
           

摘要

We examined categorical agreement between automated mycobacterial susceptibility testing methods (Mycobacterial Growth Indicator Tube [MGIT] 960 System and the VersaTREK Mycobacteria Detection and Susceptibility System) which are based on single critical concentration (CC) "breakpoints" and a commercial microbroth dilution method (Sensititre Mycobacterium tuberculosis MIC Plate [MYCOTB]) which provides an MIC value. Mycobacterium tuberculosis isolates (n=355) were tested against three first-line antimycobacterial agents (ethambutol [EMB], isoniazid [INH], rifampin [RIF]) using the MYCOTB plate and either the MGIT 960 (site 1, n=142) or VersaTREK (site 2, n=213) systems. Overall categorical agreement was 96.8%. When stratified by drug and CC-defined susceptible and resistant isolates, concordance ranged from 75% to 100%. Interpretation of MIC-based results versus established CC-based results was challenging for drugs whose CC was not represented by an exactly equivalent concentration in the manufacturer-defined dilutions on the MYCOTB plate (EMB, INH). We propose interpretations of MYCOTB plate MICs using the currently available plate configuration. (C) 2018 Elsevier Inc. All rights reserved.
机译:我们检查了自动分枝杆菌易感性测试方法(分枝杆菌生长指标管[MGIT] 960系统和versatrek分枝杆菌检测和敏感系统)之间的分类协议,这是基于单临界浓度(CC)“断点”和商业微生物稀释方法(Sensititre结核分枝杆菌晶状体板[Mycotb])提供MIC值。用Mycotb板和MgIT 960(位点1,N = 142)对三种一线抗纤维杆菌剂(乙胺醇[emb],异烟肼[inh],利福纳蛋白[RIF])测试分枝杆菌分离物(N = 355)。或Versatrek(站点2,n = 213)系统。整体分类协议为96.8%。当通过药物和CC定义敏感和抗性分离物分层时,一致性范围为75%至100%。对基于麦克酸的结果的解释与已建立的基于CC的结果是对药物的挑战,其CC在Mycotb板上的制造商定义的稀释液中的完全等同浓度而不是表示的药物。我们使用当前可用的板配置提出了对Mycotb板材MIC的解释。 (c)2018年Elsevier Inc.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号