Comparison of ABO isoagglutinin titres by three different methods: tube haemagglutination, micro-column agglutination and automated immunohematology analyzer based on erythrocyte-magnetized technology

机译:基于红细胞 - 磁化技术的三种不同方法对ABO异血素滴度的比较三种不同方法

获取原文
获取原文并翻译 | 示例
       

摘要

Background and Objectives ABO isoagglutinin titre is important for evaluating and monitoring ABO-incompatible (ABOi) stem cells or solid organ transplantations. There are several methods to measure the titre level, including the tube haemagglutination method, micro-column agglutination and erythrocyte-magnetized technology (EMT). However, few studies have reported isoagglutinin measured by EMT. Here, we compared the isoagglutinin titre of normal individuals obtained by an automated instrument with that obtained by conventional manual methods to evaluate the feasibility of replacing the manual method with the automated instrument. Materials and Methods The ABO isoagglutinin titre was measured on residual samples of healthy individuals who visited the health promotion centre of the National Cancer Center, Korea, from April to October 2015. Samples from 120 patients were collected, which included 20 males and 20 females for each blood group (A, B and O). IgM and IgG ABO isoagglutinin titres of each blood group were measured by the tube haemagglutination, micro-column agglutination and EMT techniques. The median (minimum-maximum) titres were compared, and the concordance between two methods was evaluated with the rate of results showing within one titre difference. Results The median ABO IgM and IgG titres of all blood groups obtained by the EMT method were higher than that obtained by the conventional tube haemagglutination and micro-column agglutination. Conclusion The agreement between the two methods was comparable in case of IgM but low in IgG.
机译:背景和目标ABO Isoagglutinin滴度对于评估和监测Abo-Numatible(Aboi)干细胞或固体器官移植非常重要。有几种测量滴度水平的方法,包括管血凝法,微柱凝集和红细胞磁化技术(EMT)。然而,少数研究报告了EMT测量的异血糖素。在这里,我们将通过自动化仪器获得的正常个体的异常个体滴度与通过常规手动方法获得的,以评估用自动化仪器更换手动方法的可行性。材料和方法ABO Isoagglutinin Titre是在2015年4月至10月访问韩国国家癌症中心健康促进中心的健康个体的残余样本中。收集来自120名患者的样品,其中包括20名男性和20名女性每个血液组(A,B和O)。通过管血凝集,微柱凝集和EMT技术测量每种血型的IgM和IgG Abo Isoagglutinin滴度。比较中位数(最小最小)滴度,并评估两种方法之间的一致性,并在一个滴度差异内显示结果。结果EMT方法所获得的所有血液的中位数ABO IgM和IgG滴度高于通过常规管血凝和微柱凝集获得的血液基团的滴度。结论在IgM但低于IgG的情况下,两种方法之间的协议是可比的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号