Does the Additional Articulation in Retrieved Bipolar Hemiarthroplasty Implants Decrease Trunnionosis Compared to Total Hip Arthroplasty?

机译:与总髋关节置换术相比,检索到的双极性半序塑料植入物的额外关节是否会降低耳鼻病?

获取原文
获取原文并翻译 | 示例
       

摘要

Abstract Background Trunnionosis at the modular head-neck taper interface in metal-on-polyethylene total hip arthroplasty (MoP THA) has been shown to occur, and represents a potential mode of MoP THA failure. The purpose of the present investigation is to elucidate differences in fretting and corrosion at the head-neck taper interface of prostheses retrieved from bipolar hemiarthroplasty (BH) and MoP THA. Methods A retrieval analysis of BH and MoP THA prostheses featuring a single taper design from a single manufacturer and in?vivo for a minimum 2 years was performed. Fifteen femoral heads of 28-mm diameter and corresponding femoral stems retrieved from BH were compared with MoP THA implants matched based on time in?vivo and head length (28 mm,??3 mm to 28 mm,?+8 mm). Fretting and corrosion damage scoring was completed under stereomicroscopic visualization. Results Femoral head bore tapers retrieved from BH exhibited decreased overall fretting ( P ?= .02), when compared to those retrieved from MoP THA. Total corrosion scores for all retrieved implants were positively correlated with implantation time ( ρ ?= 0.54, P Conclusion Femoral heads retrieved from BH exhibit decreased fretting damage compared to those retrieved from MoP THA. The added articulation in BH implants may decrease torque produced at the head-neck taper junction, thereby decreasing fretting. Increased fretting damage in implants from MoP THA is not associated with increased corrosion in 28-mm heads of this taper design. The longer a BH or MoP THA prosthesis is implanted, the greater the risk of damage due to corrosion.
机译:摘要出现了模块化头颈锥形界面(MOP THA)中模块化头颈锥形界面的背景耳鼻喉膜,并且代表了拖把THA失效的潜在模式。本调查的目的是阐明从双极半序塑料术(BH)和MOP THA检索的假体的头颈锥形界面处的骨折和腐蚀的差异。方法采用来自单个制造商的单一锥度设计的BH和MOP THA假肢的检索分析及至少2年。将十五股直径和从BH检索的相应股骨头的股骨头与基于时间的拖把件相匹配,搭扣(28 mm,3mm至28mm,?+ 8 mm)。在立体镜下可视化下完成烦恼和腐蚀损伤评分。结果,与从拖把Tha检索的人相比,从BH检索的股骨头孔锥体呈现出整个烦恼(p?= .02)。所有检索植入物的总腐蚀分数与植入时间呈正相关(ρα= 0.54,从BH检索的股骨头与拖把THA检索的那些相比,在BH上检索的股骨头造成的损伤减少。BH植入物中的附加关节可能会降低在生产中产生的扭矩头颈锥形连接,从而降低微动。拖把THA的植入物中的微动损伤没有与该锥形设计的28毫米头的腐蚀增加无关。植入的BH或拖把的速度越长,风险越大由于腐蚀导致的损坏。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号