首页> 外文期刊>Ultrasound in Medicine and Biology >Power Spectrum Consistency among Systems and Transducers
【24h】

Power Spectrum Consistency among Systems and Transducers

机译:系统和换能器之间的功率谱一致性

获取原文
获取原文并翻译 | 示例
           

摘要

Use of the reference phantom method for computing acoustic attenuation and backscatter is widespread. However, clinical application of these methods has been limited by the need to acquire reference phantom data. We determined that the data acquired from 11 transducers of the same model and five clinical ultrasound systems of the same model produce equivalent estimates of reference phantom power spectra. We describe that the contribution to power spectral density variance among systems and transducers equals that from speckle variance with 59 uncorrelated echo signals. Thus, when the number of uncorrelated lines of data is small, speckle variance will dominate the power spectral density estimate variance introduced by different systems and transducers. These results suggest that, at least for this particular transducer and imaging system combination, one set of reference phantom calibration data is highly representative of the average among equivalent transducers and systems that are in good working order.
机译:使用参考幻像方法来计算声学衰减和反向散射是很广泛的。然而,这些方法的临床应用受到获取参考幻像数据的限制。我们确定从相同模型的11个传感器获取的数据和相同模型的五种临床超声系统产生了参考幻像功率谱的等效估计。我们描述了对系统和换能器之间的功率谱密度方差的贡献等于具有59个不相关回声信号的散斑差异。因此,当不相关的数据线的数量很小时,散斑差异将主导由不同系统和换能器引入的功率谱密度估计方差。这些结果表明,至少对于这种特定的传感器和成像系统组合,一组参考幻影校准数据具有高度代表等效换能器和系统的平均值。