首页> 外文期刊>Physiological measurement >Monitoring the respiratory rate of preterm infants using an ultrathin film sensor embedded in the bedding: a comparative feasibility study
【24h】

Monitoring the respiratory rate of preterm infants using an ultrathin film sensor embedded in the bedding: a comparative feasibility study

机译:使用嵌入在床上用品中的超薄薄膜传感器监测早产儿的呼吸速率:比较可行性研究

获取原文
获取原文并翻译 | 示例
           

摘要

Objective: To determine the feasibility of unobtrusively monitoring the respiratory rate (RR) in preterm infants by using a film-like pressure sensor placed between the mattress and the bedding. Approach: The RR was simultaneously measured by processing the chest impedance (CI) and the ballistographic (BSG) signal acquired from the pressure sensor in 10 preterm infants of varying body weight. Nearly 27 h of data were analyzed from these infants while in different body positions including both spontaneously breathing infants and those receiving non-invasive respiratory support. Main results: The RR acquired from the BSG signal (RR-BSG) was significantly correlated (r = 0.74) to the RR derived from the CI (RR-CI) with narrow 95% limits of agreement (10 breaths min?1). A subanalysis of epochs most and least affected by infant movement yielded comparable results. Significance: Irrespective of body weight or infant position, unobtrusively monitoring the RR of preterm infants is feasible using film-like pressure sensors.
机译:目的:通过使用放置在床垫和床上用品之间的薄膜样压力传感器,确定不引人注目地监测早产儿(RR)的可行性。方法:通过处理从改变体重的10个早产婴儿的压力传感器获取的胸部阻抗(CI)和球传中(BSG)信号同时测量RR。在这些婴儿中分析了近27小时的数据,而在不同的身体位置,包括自发性呼吸婴儿和接受非侵入性呼吸载体的人。主要结果:从BSG信号(RR-BSG)中获取的RR显着相关(R = 0.74),从CI(RR-CI)衍生的RR具有窄的95%的协议限制(10呼吸分钟?1)。由婴幼儿运动最小和最小影响的时期分析产生了可比的结果。意义:无论体重还是婴儿位置,不引人注目地监测早产儿的RR使用胶片压力传感器是可行的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号