The Analog Blanking Period of Implantable Cardiac Rhythm Devices

机译:植入心律节奏装置的模拟消隐期

获取原文
获取原文并翻译 | 示例
           

摘要

Background: Analog blanking periods (BPs) that hold down the display of electrograms (EGMs) in cardiac rhythm devices have received much less attention than the well-known digital BPs which do not influence the EGM display. In Biotronik devices (Biotronik GmbH, Berlin, Germany), when a paced event initiates an analog BP in one chamber (right atrium, right ventricle [RV], or left ventricle [LV]), an identical cross-chamber analog BP starts in the other two chambers.
机译:背景技术:按住心律节奏设备中的电视图(EGMS)显示的模拟消隐时段(BPS)已经超过了不影响EGM显示器的众所周知的数字BP的关注。 在Biotronik设备(Biotronik GmbH,德国Biotronik GmbH)中,当节奏事件在一个腔室(右心房,右心室[RV]或左心室[LV])中发起模拟BP时,同样的横腔模拟BP开始 另外两个腔室。