首页> 外文期刊>Instruments and Experimental Techniques >A Nitrogen Cryostat with Adjustable Temperature and Cold Loading of Samples for the Measurement of Optical Spectra
【24h】

A Nitrogen Cryostat with Adjustable Temperature and Cold Loading of Samples for the Measurement of Optical Spectra

机译:具有可调节温度和样品冷载的氮乳酮,用于测量光谱

获取原文
获取原文并翻译 | 示例
           

摘要

A cold-loaded cryostat was designed to measure the optical spectra of thermally unstable samples. The temperature stabilization system with a resistive heater provides temperature control in the range of 80-270 K with an accuracy of +/- 0.3 K. The cryostat was used to measure the Raman spectra of molecular hydrides of silica glass at different temperatures.
机译:设计冷负载的低温恒温器以测量热不稳定样品的光谱。 具有电阻加热器的温度稳定系统在80-270k的范围内提供温度控制,精度为+/- 0.3k。低温恒温器用于测量不同温度下二氧化硅玻璃的分子氢化物的拉曼光谱。