Removing Motion Artifact From Heart Rate Signal Of Photoplethysmography Using Active Noise Cancellation

机译:使用主动噪声消除去除来自光电读物读数的心率信号的运动伪影

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

This research was aimed at reducing the error caused by motion artifact in a heart rate monitoring system to be applied in wearable devices. The heart rate signal was obtained by using the photoplethysmography (PPG) sensor. Furthermore, a 3-axis accelerometer sensor was used to sense the body motion at the PPG sensor site. The output of the accelerometer sensor may have a considerable correlation with the motion artifact in the heart rate signal. Taking the accelerometer output signal as the reference signal, an active noise cancellation (ANC) algorithm recovered the corrupted heart rate signal from the motion artifact. The performance of the ANC was evaluated using a commercial heart-rate belt as the golden standard. The heart rate detection error of our system was 3.52% with a small motion and 8.81% with a big motion. The result of our experiments proves that the ANC method is suitable for removing the motion artifact from the heart rate signal.
机译:该研究旨在减少由心率监测系统中的运动伪影引起的误差施加在可穿戴设备中。 通过使用光学溶血(PPG)传感器获得心率信号。 此外,使用3轴加速度计传感器来感测PPG传感器部位的身体运动。 加速度计传感器的输出可以与心率信号中的运动伪影具有相当大的相关性。 采用加速度计输出信号作为参考信号,有源噪声消除(ANC)算法从运动伪影中恢复了损坏的心率信号。 使用商业心率带作为金标准评估了ANC的性能。 我们系统的心率检测误差为3.52%,动作小,动作大。 我们的实验结果证明了ANC方法适合于从心率信号中去除运动伪影。