Performance measurements of a 7 mm-diameter hydrogen heat pipe

机译:直径为7毫米的氢热管的性能测量

获取原文
获取原文并翻译 | 示例

摘要

A gravity assisted heat pipe with 7-mm diameter has been developed and tested to cool a liquid hydrogen target for extracted beam experiments at COSY. The liquid flowing down from the condenser surface is separated from the vapor flowing up by a thin wall 3 mm diameter plastic tube located concentrically inside the heat pipe. The heat pipe was tested at different inclination angles with respect to the horizontal plane. The heat pipe showed good operating characteristics because of the low radiation heat load from the surroundings, low heat capacity due to the small mass, higher sensitivity to heat loads (to overcome the heat load before the complete vaporization of the liquid in the target cell) due to the higher vapor speed inside the heat pipe which transfers the heat load to the condenser.
机译:已经开发了直径为7毫米的重力辅助热管,并对其进行了测试,以冷却液氢靶,以便在COSY进行提取束实验。从冷凝器表面向下流动的液体通过一根直径3 mm的薄壁塑料管与向上流动的蒸汽分离,该塑料管同心地位于热管内部。在相对于水平面的不同倾斜角度下测试了热管。由于周围环境的辐射热负荷低,质量小导致热容量低,对热负荷的敏感性高(在目标电池中液体完全汽化之前克服了热负荷),因此热管显示出良好的工作特性。由于热管内部蒸汽速度较高,从而将热负荷传递给冷凝器。