首页> 外文期刊>Cryogenics >Development and performance test of a cryoprobe with heat transfer enhancement configuration
【24h】

Development and performance test of a cryoprobe with heat transfer enhancement configuration

机译:具有传热增强配置的冷冻探头的开发和性能测试

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

In the present study, a cryoprobe with heat transfer enhancement configuration (HTEC) is developed and its freezing performance is evaluated experimentally. Two kinds of heat transfer enhancement configurations, i.e., coiled wire insert and helical mesh insert, are proposed and used in the cryoprobe. Furthermore, in order to ensure that vapor can be discharged freely and only liquid nitrogen reaches the cryogenic section of the cryoprobe, a vapor-liquid separator is employed upstream of the cryoprobe. It is found that the precooling time of the cryoprobe is significantly shortened with the vapor-liquid separator. The enhancement of the freezing capacity with the helical mesh insert is slightly superior to that with the coiled wire insert. With the helical mesh insert, the freezing capacity of the cryoprobe can be enhanced by 41 percent, and the wall temperature can reach 111.3 K in about l0 min. Significant temperature oscillations are observed in the precooling stage of the cryoprobe with HTEC, while only slight temperature oscillations can be found without HTEC.
机译:在本研究中,开发了具有传热增强配置(HTEC)的冷冻探针,并通过实验评估了其冷冻性能。提出了两种传热增强构造,即,盘绕的钢丝插入物和螺旋网状插入物,并在冷冻探针中使用。此外,为了确保能够自由地排出蒸汽并且仅液氮到达冷冻探针的低温部,在冷冻探针的上游采用了气液分离器。已经发现,使用气液分离器可以大大缩短冷冻探头的预冷却时间。螺旋网状插入物的冷冻能力增强效果略好于盘绕线状插入物。使用螺旋网状插入件,可以将冷冻探针的冷冻能力提高41%,并且在大约10分钟内壁温可以达到111.3K。在装有HTEC的低温探头的预冷阶段,观察到了明显的温度波动,而没有HTEC的只有很小的温度波动。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号