Low wear and friction in alumina/alumina total hip joints: a hip simulator study (see comments)

机译:氧化铝/氧化铝全髋关节的低磨损和低摩擦:一项髋关节模拟器研究(请参阅评论)

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We studied the wear of state-of-the-art alumina/alumina total hip joints with the uniaxial five-station, and the three-axis single-station hip joint simulators of Helsinki University of Technology. The diameters of the joints were 26, 28 and 32 mm. Visual examination and weighing of the components showed no wear. Scanning electron microscopy showed that (a) the wear marks were slight, consisting of removal of grains, multidirectional grooving and mild abrasion, (b) the wear mechanisms were similar to those seen in components removed from patients, and (c) there was no marked difference in the specimens worn in the uniaxial vs. three-axis simulator. The coefficient of friction was measured throughout the 5 million cycle wear test with the three-axis simulator. The average value was 0.007. We conclude that alumina/alumina joints had low wear and friction under the present test conditions. The results are promising, since minimal wear is one of the prerequisites for long-term success in prosthetic joints.
机译:我们用赫尔辛基工业大学的单轴五工位和三轴单工位髋关节模拟器研究了最先进的氧化铝/氧化铝全髋关节的磨损情况。接头的直径分别为26、28和32毫米。目视检查和称量部件无磨损。扫描电子显微镜显示:(a)磨损痕迹轻微,包括去除颗粒,多向切槽和轻度磨损;(b)磨损机理与从患者身上去除的部件相似,并且(c)没有单轴与三轴模拟器中所穿样品的差异明显。使用三轴模拟器在整个500万次循环磨损测试中测量了摩擦系数。平均值为0.007。我们得出结论,在目前的测试条件下,氧化铝/氧化铝接头具有低磨损和低摩擦。结果是有希望的,因为最小的磨损是假关节长期成功的前提之一。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号