Precision of fit of implant-supported screw-retained 10-unit computer-aided-designed and computer-aided-manufactured frameworks made from zirconium dioxide and titanium: An in vitro study

机译:植入物支撑的螺杆保留的10单元的精度精度由二氧化锆和钛制成的计算机辅助设计和计算机辅助制造的框架:体外研究

获取原文
获取原文并翻译 | 示例
           

摘要

Objective: To analyze the precision of fit of implant-supported screw-retained computer-aided-designed and computer-aided-manufactured (CAD/CAM) zirconium dioxide (ZrO) frameworks. Materials and methods: Computer-aided-designed and computer-aided-manufactured ZrO frameworks (NobelProcera?) for a screw-retained 10-unit implant-supported reconstruction on six implants (FDI positions 15, 13, 11, 21, 23, 25) were fabricated using a laser (ZrO-L, N = 6) and a mechanical scanner (ZrO-M, N = 5) for digitizing the implant platform and the cuspid-supporting framework resin pattern. Laser-scanned CAD/CAM titanium (TIT-L, N = 6) and cast CoCrW-alloy frameworks (Cast, N = 5) fabricated on the same model and designed similar to the ZrO frameworks were the control. The one-screw test (implant 25 screw-retained) was applied to assess the vertical microgap between implant and framework platform with a scanning electron microscope. The mean microgap was calculated from approximal and buccal values. Statistical comparison was performed with non-parametric tests. Results: No statistically significant pairwise difference was observed between the relative effects of vertical microgap between ZrO-L (median 14 μm; 95% CI 10-26 μm), ZrO-M (18 μm; 12-27 μm) and TIT-L (15 μm; 6-18 μm), whereas the values of Cast (236 μm; 181-301 μm) were significantly higher (P 0.001) than the three CAD/CAM groups. A monotonous trend of increasing values from implant 23 to 15 was observed in all groups (ZrO-L, ZrO-M and Cast P 0.001, TIT-L P = 0.044). Conclusions: Optical and tactile scanners with CAD/CAM technology allow for the fabrication of highly accurate long-span screw-retained ZrO implant-reconstructions. Titanium frameworks showed the most consistent precision. Fit of the cast alloy frameworks was clinically inacceptable.
机译:目的:分析植入式螺杆保留计算机辅助设计和电脑辅助制造(CAD / CAM)锆(ZRO)框架的拟合精度。材料和方法:计算机辅助设计和计算机辅助制造的Zro框架(Nobelprocera?)用于六个植入物(FDI位置15,13,​​11,21,23,25)的螺钉保留的10单元植入式重建使用激光(ZrO-L,N = 6)和机械扫描仪(ZrO-M,N = 5)制造,用于数字化植入平台和囊组支撑框架树脂图案。在相同型号上制造的激光扫描的CAD / CAM钛(TIT-L,N = 6)和铸造COCRW-ALLOY框架(铸造,N = 5),并与ZRO框架类似的设计是控制。应用单螺杆测试(植入物25螺钉保留)以利用扫描电子显微镜评估植入物和框架平台之间的垂直微拍。平均微涂层由近似和颊值计算。用非参数测试进行统计比较。结果:在ZrO-L(中值14μm; 95%CI10-26μm),ZrO-M(18μm;12-27μm)和山顶之间的垂直微拍(中位数; 95%CI10-26μm)和脚跟之间没有观察到统计学上显着的成对差异。 (15μm;6-18μm),而铸造(236μm;181-301μm)的值显着高(P <0.001),而不是三个CAD /凸轮组。在所有基团(ZrO-L,ZrO-M和铸造P&LT; 0.001,滴度L P = 0.044)中观察到从植入物23至15增加植入物23至15的单调趋势。结论:具有CAD / CAM技术的光学和触觉扫描仪允许制造高精度的长跨度螺旋保留ZRO植入物重建。钛框架表现出最一致的精度。铸造合金框架的适合是临床上的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号